Sanyo MDF-793AT, MDF-593AT Alarm and safety functions, Operating Instructions, Note:

Models: MDF-593 MDF-593AT MDF-793 MDF-793AT

1 30
Download 30 pages 58.12 Kb
Page 20
Image 20
Alarm and safety functions

OPERATING INSTRUCTIONS

Alarm and safety functions

This unit has the alarm and safety functions shown in table below, and also self diagnostic function.

Table 4. Alarm and safety function

Alarm & safety

Situation

Indication

Buzzer

Safety operation

High temperature

When the chamber temp. is

 

Intermittent

 

 

tone with

Remote alarm

approx. 10oC (or 15oC) higher

Alarm lamp blinks.

alarm

than the set temp.

 

12

with 12 min. delay

 

 

min. delay

 

 

 

 

 

Power failure

At power failure

 

Intermittent

 

When power cord is

Alarm lamp blinks.

Remote alarm

alarm

tone

disconnected.

 

 

 

 

 

 

Filter check

When the condenser filter

Filter check lamp lights.

-----

-----

is clogged.

 

 

 

 

 

When there is no key

Chamber temp. is

 

Finishing of each

Auto-return

pressing in each setting

-----

displayed.

setting mode

 

mode for 90 seconds.

 

 

 

 

 

 

If the thermal sensor is

Alarm lamp blinks.

 

 

 

E01 and 50oC is

 

Remote alarm

 

disconnected.

displayed alternately.

Intermittent

 

 

Continuous

 

 

Alarm lamp blinks.

tone

 

If the thermal sensor is

operation

 

E01 and -170oC is

 

 

shirt-circuited.

displayed alternately.

 

 

 

 

 

 

 

If the cascade sensor is

Alarm lamp blinks.

 

 

 

E03 and chamber temp.

 

 

Sensor

disconnected.

 

 

is displayed alternately.

Intermittent

Remote alarm

 

abnormality

If the cascade sensor is

Alarm lamp blinks.

tone

 

 

E04 and chamber temp.

 

 

 

shirt-circuited.

 

 

 

is displayed alternately.

 

 

 

 

 

 

 

If the filter sensor is

Alarm lamp blinks.

 

 

 

E05 and chamber temp.

 

 

 

disconnected.

Intermittent

 

 

is displayed alternately.

Remote alarm

 

 

 

If the filter sensor is

Alarm lamp blinks.

tone

 

 

 

E06 and chamber temp.

 

 

 

short-circuited.

 

 

 

is displayed alternately.

 

 

 

 

 

 

Note

 

 

 

 

The filter check lamp sometimes lights at the start-up under high ambient temperature. The lamp will is off when the chamber temperature is getting lower.

Operation after power failure

The set value (setting of chamber temperature and alarm temperature) is memorized by nonvolatile memory. Accordingly, the freezer resumes the operation with setting before power failure.

19

Page 20
Image 20
Sanyo MDF-793AT, MDF-593AT Alarm and safety functions, Operating Instructions, Note:, Operation after power failure