CAUTIONS FOR USAGE
The top surface of the frame sometimes becomes hot during the initial operation or when operating in a
hot environment. This is due to the hot gas circuit for preventing dew, and is not an indication of a
malfunction.
This unit is provided with the filter check lamp. If the filter check lamp is flashed, clean the condenser
filter. Refer to page 25 “Cleaning of condenser filter”. A clogged filter may cause poor cooling and
compressor trouble.
When connecting a liquid CO2 cylinder for the back-up system , consult your dealer of a neighboring
engineering shop of high pressure gas.
Leave the remote alarm switch “OFF” until the inside temperature of the freezer reaches the
appropriate level. Leave the back-up system switch “OFF” as well. After the inside temperature
reaches the appropriate level, set the needle for the warning device operation of the temperature
recorder.
When the unit is first installed, or re-started after long period of no use, the power failure alarm may not
operate, because its Cadnica batteries have completely discharged. This does not denote a malfunction.
In this case, continue to operate the unit for about 3 hours. To fully recharge the completely discharged
batteries, the unit has to be operated for about 2 days.
Cover the control panel with the control panel cover to avoid the malfunction resulting from an
accidental contact with control panel (such as power switch OFF).