70 MI.20xx Manual
Channel Trigger Trigger modes and appendant registers
Channel window trigger for entering signals
The analog input is continuously sampled with the selected
sample rate. The upper and the lower level define a win-
dow. Every time the signal enters the the window from the
outside, a triggerevent will be detected.
Channel window trigger for leaving signals
The analog input is continuously sampled with the selected
sample rate. The upper and the lower level define a win-
dow. Every time the signal leaves the the window from the
inside, a triggerevent will be detected.
Register Value Direction set to Value
SPC_TRIGGERMODE 40000 r/w TM_CHANNEL 20040
SPC_TRIGGERMODE0 40200 r/w TM_CHXWINENTER 10040
SPC_HIGHLEVEL0 42000 r/w Sets the window’s upper level relatively to the channel’s input range. board dependant
SPC_LOWLEVEL0 42100 r/w Sets the window’s lower level relatively to the channel’s input range. board dependant
Register Value Direction set to Value
SPC_TRIGGERMODE 40000 r/w TM_CHANNEL 20040
SPC_TRIGGERMODE0 40200 r/w TM_CHXWINLEAVE 10050
SPC_HIGHLEVEL0 42000 r/w Sets the window’s upper level relatively to the channel’s input range. board dependant
SPC_LOWLEVEL0 42100 r/w Sets the window’s lower level relatively to the channel’s input range. board dependant