S12Q TROUBLESHOOTING GUIDE

SYSTEM STATUS

 

 

 

 

 

 

LAMP

 

 

 

REMARKS
STATUS
AUDIBLE

 

UV LAMP

 

 

(GREEN
ALARM

 

 

 

LED)

 

 

 

 

 

 

 

 

 

ON

OFF

 

ON

• Correct operating conditions, unit is functioning properly.

 

 

 

 

 

 

 

 

 

• The UV lamp is spent, requires replacement lamp.

 

 

 

 

 

 

 

 

 

• UV lamp not connected to power source. Check

OFF

ON

 

OFF

connection and reconnect lamp.

 

 

 

 

• Ballast has switched off. To reset ballast remove power

 

 

 

 

to unit by disconnecting power cord from electrical plug

 

 

 

 

for a minimum of 30 seconds then reapply power.

 

 

 

 

 

OFF

OFF

 

ON

• LED indicator burnt out or wire lead broken. Replace

 

LED assembly.

 

 

 

 

 

 

 

 

 

S24Q & S40Q TROUBLESHOOTING GUIDE

 

SYSTEM STATUS

 

 

 

 

 

 

 

 

 

POWER

LAMP

 

 

 

REMARKS

ON

STATUS

AUDIBLE

UV LAMP

 

 

 

(RED

(GREEN

ALARM

 

 

 

LED)

LED)

 

 

 

 

 

 

 

 

 

 

ON

ON

OFF

ON

Correct operating conditions, unit is functioning properly.

 

 

 

 

 

 

 

 

 

 

The unit is not plugged into the electrical outlet. Plug

 

 

 

 

 

unit into receptacle and ensure proper power source.

 

 

 

 

 

 

OFF

OFF

OFF

OFF

Power not connected to internal components (circuit

 

board or ballast). Reconnect power connector.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

The GFI or one of the circuit breakers is tripped. Re-set

 

 

 

 

 

the GFI or circuit breaker.

 

 

 

 

 

 

 

 

 

 

The UV lamp is spent, requires replacement lamp.

 

 

 

 

 

 

 

 

 

 

UV lamp not connected to power source. Check

ON

OFF

ON

OFF

 

connection and reconnect lamp.

 

 

 

 

 

 

Ballast has switched off. To reset ballast remove power

 

 

 

 

 

to unit by disconnecting power cord from electrical plug

 

 

 

 

 

for a minimum of 30 seconds then reapply power.

 

 

 

 

 

 

ON

OFF

OFF

ON

LED indicator burnt out or circuit board defective. Replace

 

circuit board.

 

 

 

 

 

 

 

 

 

 

 

ON

OFF

ON

ON

Defective circuit board. Contact authorized dealer.

 

 

 

 

 

 