Batteries and charging

At that point the battery is about 85% - 90% charged. It is essential to provide a full recharge at the first opportunity.

5.Do not leave the charger connected to the battery with the mains disconnected or switched off. This could result in damage to your battery being caused by deep discharge over a period of time.

6.Always switch off at the mains before disconnecting the batteries.

Safety and caution notes

1.The charger is designed for indoor use. Do not use outdoors or expose to rain, snow, spray or moisture.

2.Use only with Sonnenschein dryfit Gel type batteries of the A200, A500 or A500C groups when employed in cyclic mode.

3.The charger may be used with other brands of Gel type batteries, subject to written confirmation from the Technical Department of Sunrise Medical.

4.Do not smoke, or create sparks or flames in the vicinity of the batteries.

5.Avoid standing the charger on carpets or rugs during use and do not cover the ventilation slots of the charger or allow objects to rest on the case.

6.Do not attempt to recharge other types of batteries such as car batteries, non-rechargeable

batteries, AGM sealed batteries, wet electrolyte batteries or batteries of the wrong voltage.

7.Repairs should only be carried out by Sunrise Medical.

8.The charger is designed for mains input voltages of 220V AC to 240V AC, for chargers to operate with 110V AC (in countries such as USA or North America), a different charger will be needed, please consult local Sunrise Medical Dealer.

9.Use of an attachment, not recommended or sold by Sunrise Medical, may result in a risk of fire, electric shock or injury to persons. To reduce the risk of damage to electric plug and cord, pull by the plug rather than the cord when disconnecting the charger. Make sure the cord is located so that it will not be stepped on, tripped over, or otherwise subjected to damage or stress. An extension cord should not be used unless absolutely necessary. Use of an improper extension cord could result in a risk of fire and electric shock. If any extension cord must be used, make sure the pins on the plug of the extension cord are the same number, size and shape as those of the plug on the charger; and that the extension cord is properly wired and in good electrical condition.