Sunrise Medical Revolution instruction manual Footrest, Wheel Locks

Models: Revolution

1 23
Download 23 pages 26.17 Kb
Page 18
Image 18

X . S E T U P & A D J U S T M E N T

2.Inside or Outside Placement

a)Remove caster housing casters.

b)Remove the two eccentric bolts from

each caster.

c)Move left-side caster to right side.

d)Move right-side caster to left side.

e)Remount casters to caster bracket.

f)Tighten the eccentric bolts.

g)Replace caster housing covers.

K.FOOTREST

Use a torque of 100 in.-lbs. when

adjusting footplate.

1. Height Adjustment

a) Remove bolts on both sides of footrest and loosen footplate.

b) Remove footrest tubes and slide plug up or down inside tube to correct position.

c)The end of the plug is tapped so a 3" - 1/4"x20 bolt (included with chair) may be used to help position the plug at the correct hole.

d)Align predrilled holes in footrest and reinsert bolts.Tighten.

2.Footplate Angle Adjustment

a)Loosen flathead screws.

b)Reposition footplate to desired angle and retighten screws.

L.WHEEL LOCKS

Quickie Revolution wheelchairs are shipped with one of two types of high- mount wheel locks.Wheel locks are installed at the factory unless you have requested otherwise.

Use a torque setting of 100 in.-lbs. when setting up wheel locks.

1. High-Mount Scissor Wheel Locks:

a) Loosen the screws on the top of each clamp.

b)Slide assembly toward rear wheel until clamp embeds into tire to prevent wheel movement, when in locked position.

c)Adjust angle position.

d)Tighten screws.

X . S E T U P & A D J U S T M E N T

M. ANTI-TIP TUBES (OPTIONAL)

Quickie Designs recommends anti-tip tubes for all wheelchairs.

Use a torque setting of 100 in.-lbs. when setting up the anti-tip tubes.

1. Inserting Anti-TipTubes Into Receiver

a) Press the rear anti-tip release lever on the anti-tip tube so that both release pins are drawn inside.

b)Insert into the anti-tip tube receiver.

c)Turn the anti-tip tube down until release pin is through the receiver mounting hole.

d)Insert second anti-tip tube the same way.

2.Adjusting Anti-Tip Receiver

a)Remove anti-tip tubes from anti-tip receivers.

b)Remove button head bolt from anti-tip tube receiver.

c)Reposition bracket to desired height.

d)Reinsert button-head bolt and tighten.

e)Repeat with second anti-tip tube receiver. Both brackets should be at exactly the same height.

f)Press the rear anti-tip release lever on the anti-tip tube, so that both release pins are drawn inside.

g)Insert it into the anti-tip tube receiver.

h)Turn the anti-tip tube down until release pin is through the receiver mounting hole.

i)Insert second anti-tip tube the same way.

3.Adjusting Anti-Tip Tube Wheel The anti-tip tube wheels may have to be raised or lowered to achieve proper clearance 1 1/2" to 2".

a)Press the anti-tip wheel release button so that both release pins are drawn inside.

b)Raise or lower to one of the three predrilled holes.

c)Release button.

d)Adjust the second anti-tip tube wheel the same way. Both wheels should be at exactly the same height.

930354 Rev. B

34

35

930354 Rev. B

 

 

Page 18
Image 18
Sunrise Medical Revolution instruction manual Footrest, Wheel Locks