EDIT Menu

￿Clock Settings. Select using the general purpose knob. Then press the VALUE button on the front panel. The menu shown in Figure 3 -131 will appear.

Figure 3 ￿131: Clock Setting Menu

Set the clock source and the clock frequency by selecting the appropriate items in the side menu. The settings for clock source and clock frequency are used for both channel 1 and channel 2.

Internal Clock

Selects the internal clock. Use the numeric keys or

 

the general purpose knob to set the internal clock

 

frequency.

External Clock

Selects the external clock. The external clock signal

 

is input through the CLOCK IN connector on the

 

rear panel of the instrument.

Default Value

Sets the internal clock frequency to the default value

 

of 100.0 MHz.

 

After setting the clock source and clock frequency,

 

select O.K. from the side menu. The values for

AWG2021 User Manual

3 ￿181