Teledyne 2002 instruction manual Relay Interface Option Board, Relay Board Specifications

Models: 2002

1 37
Download 37 pages 47.27 Kb
Page 27
Image 27
7. Relay Interface Option Board

7. Relay Interface Option Board

The relay interface board is available as an option for the Model 2002. Additionally, relays may be added to either the Digital option Board (RS-232 or RS-485) or to the Analog option Board (0-10 Volts or 4-20mA). In either case the relays operate independently of the other options. The relay board allows the TTL logic setpoints (High, Low) to directly control two SPDT relays. Whenever the setpoint indicator light in the front panel is lit, the corresponding relay is energized. The relay contacts are rated for a maximum switching current of 250VA (resistive load). Connection to the relay contacts is made via connector mounted through the rear panel of the Model 2002. Screw terminals, on the mating connector, are designed for wire from 28 AWG through 14 AWG.

7.1.Relay Board Specifications

Each relay board consists of 2 Form C (SPDT) relays with the contacts wired in parallel.

Max. Switching power

120W, 250VA

(2A @ 125VAC, 1A @ 250VAC)

 

Max. Switching voltage

220VDC, 250VAC

Max. Switching current

4A DC, AC

Max. Carrying current

6A DC, AC

UL/CSA rating

1.2A @ 125VAC

...................................................................................................................................................

1.2A @ 110VDC

.....................................................................................................................................................

4.0A @ 30VDC

7.2.Relay Connector Pin Assignment

Model 2002 Vacuum Gauge

Page 27 of 37

Page 27
Image 27
Teledyne 2002 instruction manual Relay Interface Option Board, Relay Board Specifications, Relay Connector Pin Assignment