Model 200AU-NOy Option Manual Addendum
SEL END RETURNS TO SV. SET TO 315 (315oC SET VALUE)
NOTE:
DO NOT SET THE TEMPERATURE (SV) HIGHER THAN 320oC.
SECONDARY MENU (The fixed characteristics of the system).
PRESS DISPLAY ACTION
SEL HOLD TILL p-n1 SET TO 0 – REVERSE ACTION
SEL HOLD TILL p-n2 SET TO 2 = TYPE J THERMOCOUPLE
SEL p-dF DIGITAL FILTER SET TO 5
SEL P-SL LOWER LIMIT DEFAULT = 32
SEL P-SU UPPER LIMIT SET TO 350 (350ºC)
SEL P-Ab ALARM NONE
SEL P-An HYTERESIS NONE
SEL P-dP PROG DECIMAL SET TO 0
SEL P-48 COLD JUNCTION SET TO 2 (TYPE J)
SEL r5 RAMP TO SET POINT LEAVE AT 0
SEL PVOF PROCESS OFFSET LEAVE AT 0
SEL SVOF SET POINT OFFS LEAVE AT 0
SEL P-F SET TO 0 (0 = oC, 1 = oF)
SEL P-n1
SEL HOLD TILL P
SEL SCROLL TO A7 SET TO 1 (AUTOTUNE )
SV SET VALUE SHOULD BE 315°C
02808 Rev. E Page 34