Advanced FEATURES of the M400E analyzer M400E Ozone Analyzer Operator’s Manual

7.1.5.9. The Compact Report Feature

When enabled, this option avoids unnecessary line breaks on all RS-232 reports. Instead of reporting each
parameter in one channel on a separate line, up to five parameters are reported in one line.
The COMPACT DATA REPORT generally cannot be accessed from the standard iDASfront panel menu, but is
available via the instruments communication ports by using APICOM or the analyzer’s standard serial data
protocol.

7.1.5.10. The Starting Date Feature

This option allows the user to specify a starting date for any given channel in case the user wants to start data
acquisition only after a certain time and date. If the STARTING DATE is in the past (the default condition), the
iDAS ignores this setting and begins recording data as defined by the REPORT PERIOD setting.
The STARTING DATE generally cannot be accessed from the standard iDAS front panel menu, but is available
via the instruments communication ports by using APICOM or the analyzer’s standard serial data protocol.
7.1.6. DISABLING/ENABLING DATA CHANNELS
Data channels can be temporarily disabled, which can reduce the read/write wear on the disk-on-chip.
To disable a data channel, follow the instruction shown in Section 7.1.5 then press:
Starting at the EDIT CHANNEL MENU
SETUP X.X 0) CONC: ATIMER 1, 800
PREV MEXT INS DEL EDIT PRNT EXIT
SETUP X.X NAME: CONC
<SET SET> EDIT PRNT EXIT
Continue pressing <SET or SET> until ...
SETUP X.X CHANNEL ENABLE:ON
<SET SET>EDIT PRNT EXIT
Use the PREV and
NEXT keys to scroll to
the DATA CHANNEL
to be edited
SETUP X.X CHANNEL ENABLE:ON
ON ENTR EXIT EXIT discards the new
setting
ENTR accepts the
new setting
Toggle these keys to
enable or disable the
CHANNEL

92 04315 Rev. C1