Teledyne DNV-33D instruction manual Operating Principle, 1.0SPECIFICATIONS, Section

Models: DNV-33D

1 12
Download 12 pages 15.19 Kb
Page 7
Image 7
Operating Principle

SECTION II

Operating Principle

The operation of Hastings vacuum gauges uses a patented noble metal thermopile circuit. The hot junctions of the thermopile are heated directly by an AC current while an equal number of cold junctions are kept at ambient temperature by heavy mounting studs. Thus a DC voltage is then generated between the hot and cold junctions. As the pressure decreases, the lowering of the thermal conductivity of the gas surrounding the hot junction tends to increase the tempera- ture of the hot junctions, thus increasing the output of the thermopile. This change in output is then amplified, linearized, and calibrated as a function of pressure.

1.0SPECIFICATIONS

Indicator Model

DNV-33D

Case Dimension

Front Panel 96mm SQ. Depth 10 ¾”

Gauge Tube Type

DV-33D-1, DV-33D-2, DV-33D-3*

Gauge Tube Shell

Welded stainless steel shell

Overpressure

1500 PSIG max.

Panel Cut-out

3-1/16” (77.5 mm) H x 3-5/8” (92mm)W

Range

1-1000 mTorr

Readout

Digital Meter

Power

115 VAC, less than 6 watts

Outputs

Linear 0-1 VDC analog @ 4 ma. Max.

Calibration

Calibrated for air or nitrogen

*Dash number Defines System Fitting Differences:

-1

Weld Fitting

 

-2

Cajon #SS-4 VCR Male Gland

 

-3

Cajon #SS-4 VCR Female Gland

 

page 7

Page 7
Image 7