Ultra Sensitivity Model M200EU NO/NO2/NOx Analyzer

Addendum to M200E

2.SPECIFICATIONS, AGENCY APPROVALS, AND WARRANTY

Min/Max Range (Physical Analog Output)

Measurement Units

Noise at Zero1

Noise at Span1

Lower Detectable Limit1

Zero Drift2

Span Drift2

Linearity

Precision

Lag Time1

Rise / Fall Time1, 3

Sample Flow Rate

Temperature Range

Humidity

Hydrocarbon Interference

Voltage Coefficient

Dimensions H x W x D

Weight, Analyzer

Weight, Ext Pump Pack

Power, Analyzer

Power, Analyzer4

Power, Ext Pump

Power, Ext Pump4

Analog Outputs

Analog Output Ranges

Analog Output Resolution

Status Outputs

Control Inputs

Min: 0-5 ppb of Full Scale Max 0-2000 ppb of Full Scale

ppb, ug/m3 (user selectable Single, independent or auto ranges) <25 ppt RMS

<0.25% RMS of reading above 50 ppb 50 ppt RMS

<0.1 ppb / 24 hours <0.2 ppb / 7 days

<0.5% of reading or 50 ppt RMS whichever is greater / 24 hrs <1% of reading or 100 ppt RMS whichever is greater / 7 days

1% of full scale or 0.1 ppb whichever is greater 0.5% of reading above 50 ppb

20 sec

95% in <50 sec

1000 cc/min. ± 10%

20-30oC within drift and noise specifications, and operating with EPA equivalency 5-35oC safe operating range

0-95% RH non-condensing

Propylene rejection ratio > 20,000:1 Ethylene rejection ratio > 40,000:1

< 0.1% per V

7" x 17" x 23.5" (178mm x 432 mm x 597 mm)

18 kg (40 lbs)

9.5 kg (21 lbs)

100V~ 50/60 Hz, 120V~ 60 Hz, 230V~ 50 Hz, 2.5A, 125 watts 230V~ 50 Hz, 2.5A

110V~ 60 Hz, 220V~ 50 Hz, 240V~ 50Hz, 150 watts 230V~ 50 Hz, 2.2A

Four (4) Outputs

All outputs: 100 mV, 1, 5, 10V

Three outputs convertible to 4-20 mA isolated current loop All ranges with 5% under/over range

1 part in 4096 of selected full-scale voltage

8 Status Outputs from opto-isolators, 7 defined, 1 spare 6 Control Inputs, 4 defined, 2 spare

2

TAPI 05385 Rev B

PRINTED DOCUMENTS ARE UNCONTROLLED

Page 6
Image 6