Teledyne Analytical Instruments
3 Installation Model 2010B
3-10 Part I
Table 3-4: Required RS-232 Options
Parameter Setting
Baud 2400
Byte 8 bits
Parity none
Stop Bits 1
Message Interval 2 seconds
3.3.9 Remote Probe Connector
The 2010B is a single-split configuration analyzer, the Remote Probe
connector is used to interface the Analysis Unit. See Figure 3-6.
Figure 3-6: Remote Probe Connector Pinouts
The voltage from these outputs is nominally 0 V for the OFF and
15 V dc for the ON conditions. The maximum combined current that can be
pulled from these output lines is 100 mA. (If two lines are ON at the same
time, each must be limited to 50 mA, etc.) If more current and/or a different
voltage is required, use a relay, power amplifier, or other matching circuitry
to provide the actual driving current.
In addition, each individual line has a series FET with a nominal ON
resistance of 5 ohms (9 ohms worst case). This could limit the obtainable
voltage, depending on the load impedance applied. See Figure 3-7.
Insert wire
here.
Turn cw to hold
ccw to
loos en w ire.
SAMPLE INSPAN IN ZERO IN EXH AUST
+15VSwitchingGround