Teledyne pjnmoo385 Sensor Installation/Replacement Procedure for the SNACK O2 Analyzer, Figure A

Models: pjnmoo385

1 22
Download 22 pages 54.91 Kb
Page 14
Image 14
Sensor Installation/Replacement Procedure

Sensor Installation/Replacement Procedure

for the SNACK O2 Analyzer

STEP 1

Place the analyzer face down on a soft surface and remove the six screws from the back of the case. Without opening the case, place the analyzer face up.

STEP 2

With the analyzer faced up, carefully lift the top section from the bottom section. Place the top section face down to the right of the lower section. Be aware that the two sections are connected by a non-detachable ribbon cable.

STEP 3

Locate the sensor connector cable (ribbon cable). Gently push the cable plug into the sensor socket until the two Side Ejector locks close.

STEP 4

Locate the blue sample supply tube. Gently push the tube onto the center sensor barb.

STEP 5

Locate the Sample Discharge tube with the lure lock fitting. Screw the lure lock to the sensor discharge fit- ting.

STEP 6

Carefully place the sensor back down into the case as shown. Close the case and replace the six screws.

CALIBRATE THE SENSOR BE- FORE USE.

Figure A

Teledyne Analytical Instruments

14

Page 14
Image 14
Teledyne pjnmoo385 owner manual Sensor Installation/Replacement Procedure for the SNACK O2 Analyzer, Figure A