4.6.3.6. AIN Calibration
This is the sub-menu to conduct the analog input calibration. This calibration should only
be necessary after major repair such as a replacement of CPU, motherboard or power
supplies. Activate the ANALOG I/O CONFIGURATION MENU (see Section 4.6.3),
then press:
Exit at any tim e to
ret urn to the ma in
DIAG menu
DIAG
A
NALOG I / O CONFIGURATION
PREV NEXT ENTR EXIT
Cont inue pres sing SE T> until …
DIAG AIO
A
IN CALIB RA T ED: NO
< SET SET> CAL EXIT
DIAG AIO
A
IN CAL IBRAT ED: YES
< SET SET> CAL EXIT
Instrument
calibrates
automatically
Exit to return to the
ANALOG I/O
CONFIGURATION
MENU
STARTING FROM ANALOG I / O CONFIGURATIONMENU
DIAG AIO CALIBRATIN G A/D ZERO
CALIBRATING A/D SPAN