Teledyne API - Model T200H/T200M Operation Manual Operating Instructions
135
5. Set the reporting range scale for the data type chosen.
6. Set the update rate for the channel.
7. Calibrating the output channel. This can be done automatically or manually for
each channel (see Sections 4.13.6).
To access the analog I/O configuration sub menu, press:
AIO Configuration Submenu
SAMPLE A1:NXCNC1=100PPM NOX=XXX.X
< TST TST > CA L SETUP
SETUP X.X PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK MORE EXIT
SETUP X.X SECONDARY SETUP MENU
COMM VARS DIAG ALRM EXIT
SETUP X.X ENTER PASSWORD:818
8 1 8 ENTREXIT
DIAG SIGNA L I/O
NEXT ENTR EXIT
Continue pressing NEXT until ...
DIAG ANALOG I/O CONFIGURATION
PREV NEXT ENTR EXIT
DIAG AIO A OUTS CALIBRATED: NO
<SET SET> CAL EXIT
DIAG AIO DATA_OUT_1: 5V, NXCNC1, NOCAL
<SET SET> EDIT EXIT
DIAG AIO AIN CALIBRATED: NO
<SET SET> CAL EXIT
DIAG AIO DATA_OUT_2: 5V, NXCNC1, NOCAL
<SET SET> EDIT EXIT
DIAG AIO DATA_OUT_3: 5V, NXCNC1, NOCAL
<SET SET> EDIT EXIT
DIAG AIO DATA_OUT_4: 5V, NXCNC1, NOCAL
<SET SET> EDIT EXIT
07270B DCN6512