Teledyne API - Model T200H/T200M Operation Manual Getting Started
67
3.7.2.2. O2 Calibration Method
STEP 1 – Set O2 span gas concentration :
Set the expected O2 span gas concentration.
This should be equal to the percent concentration of the O2 span gas of the selected
reporting range (default factory setting = 20.8%; the approximate O2 content of ambient
air).
SAMPLE GAS TO CAL:NOX
NOX O2 ENTREXIT
SAMPLE A1:NXCNC1=100PPM NOX=XXX.X
< TST TST > CAL SETUP
M-P CAL A1:NXCNC1 =100PPM NOX=X.XXX
<TST TST> ZERO SPAN CONC EXIT
SAMPLE GAS TO CAL:O2
NOX O2 ENTREXIT
M-P CAL O2 SPAN CONC:20.8%
020.80 ENTREXIT
The O2span concentration value automatically defaults to
20.8 %.
If this is not the the concentration of the span gas being
used, toggle these buttons to set the correct concentration
of the O2calibration gases.
EXIT ignores the new
setting and returns to
the previous display.
ENTR accepts the new
setting and returns to
the previous menu.