Getting Started Teledyne API - Model T200H/T200M Operation Manual
66
3.7.2. BASIC O2 SENSOR CALIBRATION PROCEDURE
If your instrument has an O2 sensor option installed that should be calibrated as well.
3.7.2.1. O2 Calibration Setup
The pneumatic connections for calibrating are as follows:
Calibrated N2
at HIGH Span
Concentration
Calibrated O2
at 20.8% Span
Concentration
Source of
SAMPLE GAS
Removed during
calibration
Instrument
Chassis
SAMPLE
EXHAUST
PUMP
VENT
3-way
Valve
Manual
Control Valve
VENT here if input
is pressurized
Figure 3-25: O2 Sensor Calibration Set Up
O2 SENSOR ZERO GAS: Teledyne API’ recommends using pure N2 when calibration
the zero point of your O2 sensor option.
O2 SENSOR SPAN GAS: Teledyne API’ recommends using 21% O2 in N2 when
calibration the span point of your O2 sensor option.