Getting Started Teledyne API - Model T200H/T200M Operation Manual
68
STEP 2 – Activate O2 sensor stability function
To change the stability test function from NOx concentration to the O2 sensor output,
press:
SETUP X.X STABIL_GAS:O2
NO NO2 NOX O2 ENTR EXIT
Press ENTR to keep
changes, then press
EXIT 3 times to return
to SAMPLE menu
SAMPLE A1:NXCNC1=100PPM NOX=XXX.X
< TST TST > CAL SETUP
SETUP X.X PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK MORE EXIT
SETUP X.X SECONDARY SETUP MENU
COMM VARS DIAG ALRM EXIT
SETUP X.X ENTER PASSWORD:818
8 1 8 ENTREXIT
SETUP X.X 0) DAS_HOLD_OFF=15.0 Minutes
<PREV NEXT> JUMP EDIT PRNT EXIT
SETUP X.X 2) STABIL_GAS=NOX
<PREV NEXT> JUMP EDIT PRNT EXIT
SETUP X.X STABIL_GAS:NOX
NO NO2 NOX O2 ENTR EXIT
Continue pressing NEXT until ...
Note Use the same procedure to reset the STB test function to NOx when the O2
calibration procedure is complete.