Teledyne API - Model T200H/T200M Operation Manual Operating Instructions
77
4.3. CALIBRATION MODE
4.3.1. CALIBRATION FUNCTIONS
Pressing the CAL key switches the T200H/M into calibration mode. In this mode, the
user can calibrate the instrument with the use of calibrated zero or span gases.
If the instrument includes the zero/span valve option, the display will also include
CALZ and CALS buttons. Pressing either of these buttons also puts the instrument into
multipoint calibration mode.
The CALZ button is used to initiate a calibration of the zero point.
The CALS button is used to calibrate the span point of the analyzer. It is
recommended that this span calibration is performed at 90% of full scale of the
analyzer’s currently selected reporting range.
Because of their critical importance and complexity, calibration operations are described
in detail in Section 5.
4.4. SETUP MODE
The SETUP mode contains a variety of choices that are used to configure the analyzer’s
hardware and software features, perform diagnostic procedures, gather information on
the instruments performance and configure or access data from the internal data
acquisition system (DAS). The areas access under the Setup mode are:
Table 4-4: Primary Setup Mode Features and Functions
MODE OR FEATURE MENU
BUTTON DESCRIPTION
Analyzer Configuration CFG Lists key hardware and software configuration information
Auto Cal Feature ACAL Used to set up an operate the AutoCal feature. Only appears if
the analyzer has one of the internal valve options installed
Internal Data Acquisition
(DAS) DAS Used to set up the DAS system and view recorded data
Analog Output Reporting
Range Configuration RNGE Used to set the units of measure for the display and set the
dilution ratio on instruments with that option active.
Calibration Password Security PASS Turns the password feature ON/OFF
Internal Clock Configuration CLK Used to Set or adjust the instrument’s internal clock
Advanced SETUP features MORE This button accesses the instruments secondary setup menu
07270B DCN6512