Model T200U-NOy Option Manual Addendum

If you have a leak checker:

1.Turn off power to the analyzer and 501Y, and unplug the external Pump.

2.Disconnect the external Pump from the analyzer’s rear panel. Cap the sample inlet of the external Converter, the BYPASS OUT port of the 501Y, the ozone generator air inlet (located inside the analyzer, Figure 6), and connect the leak checker to the instrument exhaust port.

CAUTION!

Pressure must be less than 15 PSI.

3.Pressurize the system and check for leaks by watching overall pressure. The pressure should not drop more than 1"Hg (.5psi) in 5 minutes.

If the instrument fails the pressure test, each fitting needs to be leak checked to find the location.

NOTE

Be careful that the system is always pressurized so as not to draw soap solution into the plumbing system. Make sure you dry off any accumulated bubble solution. Start by checking the external fittings first.

30

07303C DCN6646

Page 32
Image 32
Teledyne T200U-NOy manual Pressure must be less than 15 PSI

T200U-NOy specifications

The Teledyne T200U-NOy is an advanced gas analyzer specifically designed to measure nitrogen oxides (NOx) concentrations in a variety of applications, including industrial emissions monitoring, environmental assessments, and combustion control. It builds upon Teledyne's legacy of producing high-quality analytical instruments, making it a reliable choice for professionals in the field.

One of the main features of the T200U-NOy is its ability to measure multiple nitrogen oxides, including nitrogen dioxide (NO2), nitric oxide (NO), and nitrous oxide (N2O), which are critical for understanding and managing air quality and combustion efficiency. The instrument offers high sensitivity and accuracy, allowing for real-time monitoring of NOx levels with a detection limit that suits stringent regulatory requirements.

The T200U-NOy utilizes advanced chemiluminescence technology for NO measurement, which is widely recognized for its reliability and precision. This method involves the reaction of NO with ozone (O3) to produce excited NO2, which subsequently releases light as it returns to a lower energy state. The intensity of this light is measured and directly correlates to the concentration of NO in the sample, providing highly accurate readings.

In addition to its core measurement capabilities, the T200U-NOy is equipped with a variety of features that enhance its functionality. Its user-friendly interface allows for easy configuration and data retrieval, making it accessible to a wide range of users, from seasoned professionals to those new to gas analysis. The device supports multiple communication protocols, including analog and digital outputs, facilitating integration into existing monitoring systems.

The T200U-NOy is constructed with robust materials designed to withstand harsh industrial environments, ensuring longevity and reliability. It also includes features for easy maintenance, such as user-replaceable parts, which can significantly reduce downtime.

Overall, the Teledyne T200U-NOy stands out in the field of gas analyzers, combining cutting-edge measurement technologies with user-centric features and robust construction. Its comprehensive capabilities make it an essential tool for accurately monitoring nitrogen oxides, ultimately contributing to better environmental protection and compliance with air quality regulations.