RT318

How to wire

Shielded twisted cable

Internal circuit side

 

 

 

 

nA

 

RTD

 

 

 

nB

 

(Pt100)

 

 

 

nC

 

 

 

 

 

AG

 

 

 

 

nG

RTD

:

Resistance

 

 

FG

Temperature

Detector

 

 

n

:

Channel

number

 

 

AG

:

Analog

ground

 

 

FG : Frame ground

 

 

 

If measurement values are unstable due to noise influences, change the grounding for the shield in the following order for stable measurement.

(1)

RTD

(Pt100)

(2)

RTD

(Pt100)

(3)

RTD

(Pt100)

Shielded twisted cable

Internal circuit side

nA

 

nB

 

nC

AG

nG

FG

 

Shielded twisted cable

Internal circuit side

 

nA

 

 

nB

 

 

nC

AG

 

nG

 

FG

 

 

Shielded twisted cable

Internal circuit side

nA

 

nB

 

nC

AG

nG

FG

 

Supplementary

To minimize the influences of the wire resistance, be sure the wire length to be same of each cable from the terminal nA , nB or nC to RTD.

4

Page 4
Image 4
Toshiba RT318 manual How to wire