Resistant to electrostatic discharge

Supports loads up to 700VA/450 watts

36 minutes half load runtime (with 350VA load), 14.5 minutes full load runtime (with 700VA load)

6 protected hospital-grade outlets; IEC C14 inlet on 6-inch dongle

RS-232 and USB monitoring ports for remote monitoring and unattended shutdown

PowerAlert unattended shutdown software available via free download from www.tripplite.com (cabling included)

Complete surge protection, line isolation and EMI/RFI noise filtering

2 year product warranty

5 diagnostic LEDs indicate AC present, on battery, voltage correction, overload & battery low/replace

Battery replacement door Grounding lug

Specifications

SYSTEM OVERVIEW

Voltage compatibility

230V AC

 

 

Frequency

50/60 Hz (auto-select)

compatibility

 

 

 

OUTPUT

 

Output VA

700

 

 

Output watts

450

 

 

Output nominal

230V AC

voltage

 

 

 

Output voltage

LINE MODE: Sine wave line voltage 230V (-18% / +8%)

regulation

BATTERY MODE: PWM sine wave output 230V +/-5%

 

 

Output frequency

LINE MODE: Passes line frequency of 50Hz +/-10%

regulation

BATTERY MODE: Inverter output regulated to 50/60Hz +/-0.5Hz

 

 

Outlet quantity / type

6 IEC C13

 

 

Overload protection

Resettable circuit breakers

 

 

INPUT

 

Maximum input amps

3.6A / 828 watts

 

 

Input connection type

IEC C14 inlet

 

 

Input cord length

6 inch dongle., 16 gauge

 

 

Recommended

10A 230V

Electrical Service

 

 

 

BATTERY

 

Full load runtime

Half load runtime

DC system voltage

Typical battery lifespan

14.5minutes (700VA) 36 minutes (350VA)

36V DC

3-6 years, depending on usage

Battery recharge rate

2-4 hours to 90%

 

 

VOLTAGE REGULATION

 

Voltage regulation

Line interactive UPS offers 2 boost levels of brownout correction and a single level of overvoltage adjustment to

description

maintain usable, computer grade output without consuming battery power over an input voltage range of 155 to 274V

 

AC

 

 

Overvoltage

Input voltages between 206 and 245V AC are reduced by 9%

correction

 

 

 

Direct pass through

Input voltages between 200 and 244V AC are passed through unchanged

Brownout correction

Input voltages between 183 and 199V AC are boosted by 10%

Page 2
Image 2
Tripp Lite SMX700HG warranty Specifications, System Overview, Output, Input, Battery