NOTE: According to extensive clinical examinations, the temperature readings given by an ear thermometer are equivalent to those given by its oral counterparts.
Use of this ear thermometer should not replace a visit to your doctor.
TROUBLESHOOTING
1.Consistent low temperature readings
a.The probe is not positioned properly in the ear canal. The tip of the probe must be snug and fully seated against the opening of the ear canal. Failure to properly position the probe may lead to a low temperature reading.
b.The waterproof lens is dirty. Clean the lens with a soft cloth moistened with water or alcohol.
2.The measurement is out of the range of the ear thermometer
a.Switch the ear thermometer on before positioning it in the ear. Please properly position the ear thermometer in the ear canal and then repeat the measurement.
3.Low battery warning
a.Battery needs to be replaced.
See Battery Installation section of the manual.
NOTE: If you believe your thermometer is defective or otherwise not operating properly, please check the battery first. If the thermometer is still not operating properly or if you are having any other technical problems, please contact:
Unisar Inc.
15 West 36th Street
New York, NY 10018
TECHNICAL SPECIFICATIONS
•For household use only
•Temperature readings are to the nearest 0.1°F or °C
•Temperature measurement range: 93.2° ~ 109.4°F/34°C ~ 43°C
•Ambient operating environment:
60.8~ 104°F (16° ~ 40°C)
•Ambient storage environment:
•Laboratory accuracy against an infrared reference source:
95°~102.2°F/35° ~ 39°C: +/- 0.4°F/0.2°C
less than 96.8°F/36°C: +/- 0.5°F/0.3°C
greater than 102.2°F/39°C:
•Requires one
•Equipment is not category AP/APG