Specification

OUTPUT

Voltage Range:

0V to 42V minimum

Current Range:

0A to 10A minimum

Output Voltage Setting:

By coarse and fine controls.

Output Current Setting:

By single logarithmic control.

Operating Mode:

Constant voltage or constant current with automatic cross-over.

Output Switch:

Electronic. Preset voltage and current displayed when off.

Output Terminals:

4mm terminals on 19mm (0.75”) pitch.

Output Impedance:

Typically <5min constant voltage mode.

 

Typically >20kin constant current mode.

Output Protection:

Output will withstand up to 48V forward voltage.

 

Reverse protection by diode clamp for reverse currents up to 3A.

Over-temperature

The output will be tripped off if a fault causes the internal

Protection:

temperature to rise excessively.

Load & Line Regulation:

<0.01% of maximum output for a 90% load change or 10% line

 

change.

Ripple & Noise

Typically <2mVrms, <10mV pk-pk, constant voltage mode.

(20MHz bandwidth):

 

Transient Response:

<200µs to within 50mV of set level for 90% load change.

Temperature Coefficient:

Typically <100ppm/°C

Status Indication:

Output on lamp. Constant current mode lamp.

METER SPECIFICATIONS

Meter Types:

Dual 4 digit meters with 14mm (0.56") LEDs. Reading rate 4 Hz.

Meter Resolutions:

10mV, 10mA

Meter Accuracies:

Voltage 0.3% of reading ± 2digit,

 

Current 0.6% of reading ± 2digit

GENERAL

AC Input:

Factory set for: 110V - 120V AC ±10% or 220V - 240V AC ±10%,

 

50/60Hz. Installation Category II.

Power Consumption:

800VA max.

Operating Range:

+5ºC to +40ºC, 20% to 80% RH.

Storage Range:

40ºC to + 70ºC.

Environmental:

Indoor use at altitudes up to 2000m, Pollution Degree 2.

Safety:

Complies with EN610101.

EMC:

Complies with EN61326.

Size:

140 x 160 x 320mm (WxHxD).

Weight:

3.6kg