Omega ILD24-UTP, ILD44-UTP manual Thermocouple Field Calibration Submenu

Page 39

THERMOCOUPLE FIELD CALIBRATION SUBMENU:

CAUTION: Do not perform the following steps until you fully understand this entire section.

RTD and Process are perfectly calibrated. This section is applicable to Thermocouple (TC) calibration only.

Be sure that the TC being used to calibrate the meter is of the type selected in the TC submenu. Place the TC in an ice-bath (or other

0°C / 32°F environment). In ambient temperature conditions: connect the TC to the meter, apply power to the meter.

CAUTION: Do not proceed with TC calibration unless the above conditions have been in effect for at least one hour.

Press

 

7)

Display shows CAL

°

.

a

1

Press

d

8)

Display shows flashing 0000.

Press

a *

9)

Display will still show flashing 0000.

Press

d *

10) Display shows OUT1 (meaning Calibration is complete)

*If you accidently engage the flashing 0000 (CAL° alert) simply re-press the last button you pressed, to avoid unintentionally mis-calibrating your meter.

35

Image 39
Contents User’s Guide Servicing North America Table of Contents List of Tables List of FiguresTIP Description Part IntroductionEMC Considerations Safety ConsiderationsBefore You Begin Mounting Part SetupFront Panel Annunciators Mounting iLD Big Display Through PanelMounting iLD Big Display on Bail Option PowerRear Panel Connections InputMain Power Connections Electrical Installation Power ConnectionsThermocouple Thermocouple Wiring Hookup TC Wire Color Chart3 Two/Three/Four-Wire RTD Process Current Process VoltageWithout Sensor Excitation Wiring Outputs RS-232 Output Wiring Hookup b RS-485 Output Wiring HookupOutputs Wiring Hookup 12 Excitation Output Turning your Controller On for the First Time Operation Configuration Mode IntroductionButton Function in Configuration Mode PartFlow Chart for ID and Setpoints Menu ConfigurationEntering Your NON-DEFAULT Full Security ID Number Entering Your NON-DEFAULT SETPOINT/ID Security ID NumberID Number Setpoint Set PointsConfiguration Menu Enter Configuration MenuInput Type Menu Thermocouple Types Input Type ThermocoupleEnter Input Type Menu Thermocouple SubmenuRTD Value Submenu Input Type RTDRTD Submenu RTD TypesProcess Types Reading ConfigurationInput Type Process Process SubmenuFilter Constant Submenu Enter Reading Configuration MenuTemperature Unit Submenu Decimal Point SubmenuINPUT/READING Scale and Offset Submenu Reading Configuration If Process was selectedRange Conversion Number Conversion TableEnter Alarm 1 Menu AlarmAlarm 1 ENABLE/DISABLE Submenu Alarm 1 ABSOLUTE/DEVIATION SubmenuAlarm 1 LATCH/UNLATCH Submenu Active Submenu Contact Closure SubmenuAlarm ENABLE/DISABLE AT Power on Alarm 1 LOW Value SubmenuAlarm 1 HI Value Submenu Enter Analog Output Menu Analog Output RetransmissionOUT Analog Output ENABLE/DISABLE SubmenuCURRENT/VOLTAGE Submenu ReadingFor example Enter Alarm 2 Menu Enter Loop Break Time Menu Alarm 2 ENABLE/DISABLE SubmenuLoop Break ENABLE/DISABLE Submenu Loop Break TimeReading Adjust Submenu Setpoint Deviation ENABLE/DISABLE SubmenuLoop Break Time Value Submenu Thermocouple Field Calibration Submenu Output Flow Chart for OutputEnter Output 1 Menu Self SubmenuMINIMUM/PERCENT LOW Submenu Control Type Output MAXIMUM/PERCENT High SubmenuAction Type Submenu Auto PID SubmenuAnti Integral Submenu Proportional Band Submenu Start Auto Tune PIDReset Setup Submenu Rate Setup SubmenuCycle Time Submenu Deadband Submenu Damping Factor SubmenuControl Type Submenu Enter Output 2 MenuAction Type Submenu Cycle Time Submenu Enter Ramp and Soak Menu Ramp ENABLE/DISABLE SubmenuSoak ENABLE/DISABLE Submenu Ramp & SoakSoak Value Submenu Ramp Value SubmenuEntering or Changing Your NON-DEFAULT ID Code ID CodeEnter ID Code Menu Entering or Changing Your Default ID Code Full Security Level SubmenuSETPOINT/ID Security Level Submenu Communication Option 13 Flow Chart for Communication OptionParity Submenu Enter Communication Option MenuCommunication Parameters Submenu Baud SubmenuLine Feed Submenu Stop BIT SubmenuBUS Format Submenu Modbus Protocol SubmenuData Flow Mode Submenu Communication Interface Standard SubmenuData Separation Character Submenu Peak Value Submenu Alarm Status SubmenuData Format Submenu Main Reading SubmenuTransmit Time Interval Value Submenu Address Setup SubmenuAddress Value Submenu Transmit Time Interval SubmenuNormal Color Display Submenu Alarm 1 Display Color SubmenuDisplay Color Selection Enter Display Color Selection MenuGreen RED Amber Alarm 2 Display Color SubmenuExample Display Colors change sequencesRED Green Control Part SpecificationsNmrr CmrrGeneral Control Output 1SSR InsulationInput Type Range Accuracy Input PropertiesFactory preset value Part Factory Preset ValuesMenu Items Factory Preset Values Notes CE Approval Information Page Return REQUESTS/INQUIRIES WARRANTY/DISCLAIMERData Acquisition TemperaturePRESSURE, Strain and Force FLOW/LEVEL
Related manuals
Manual 2 pages 47.34 Kb

ILD24-UTP, ILD44-UTP specifications

The Omega ILD44-UTP and ILD24-UTP are advanced intrusion detection sensors designed for versatile security applications. These devices stand out due to their cutting-edge technology, ensuring reliability and high performance in safeguarding various environments.

One of the primary features of the ILD44-UTP and ILD24-UTP is their unique dual-tech detection capability. By utilizing both passive infrared (PIR) and microwave detection, these sensors significantly reduce false alarms while ensuring accurate intrusion detection. The combination of these technologies allows for effective monitoring in different settings, including residential, commercial, and industrial spaces.

The ILD44-UTP model is equipped with a wider detection range compared to its counterpart, the ILD24-UTP. It is capable of covering a detection area of up to 44 meters, making it ideal for larger spaces. Conversely, the ILD24-UTP provides effective coverage for smaller areas, reaching up to 24 meters. This range flexibility gives users the option to select a model that best fits their particular needs.

Another important characteristic is the sensor's weather-resistant design, which allows it to be installed outdoors without compromising performance. The ILD44-UTP and ILD24-UTP are rated for various environmental conditions, ensuring they remain operational in rain, snow, or extreme temperatures. This durability makes them suitable for use on building perimeters, parking lots, and other outdoor security applications.

Both models are equipped with advanced signal processing algorithms that enhance their ability to differentiate between real threats and environmental disturbances, greatly minimizing false alarm rates. This feature is especially critical in settings where frequent alarms can lead to alarm fatigue among security personnel.

The installation process for both devices is straightforward, with user-friendly mounting options. They can be easily integrated into existing security systems, offering flexibility for users to combine them with other security measures such as cameras and access control systems.

Additionally, the ILD44-UTP and ILD24-UTP come with various output options, including relay and voltage outputs, enabling seamless integration with alarm systems and monitoring solutions. Their compatibility with different protocols ensures they can be tied into a wide array of security infrastructures.

Overall, the Omega ILD44-UTP and ILD24-UTP sensors represent a significant advancement in intrusion detection technology, combining robust performance with innovative features that cater to a variety of security needs. Their versatility, durability, and reliability position them as top choices for enhancing security across numerous applications.