Fluke 51 & 52 Series II user manual Measuring Temperatures, Connecting a Thermocouple, Type Color

Page 15

51 & 52 Series II

Measuring Temperatures

Measuring Temperatures

Notes

Connecting a Thermocouple

The display shows "- - - -" when a thermocouple

is not connected.

 

To change the thermocouple type, see “Changing Setup Options.” The North American ANSI Color Code is:

Type

J

K

E

T

N

 

 

 

 

 

 

Color

Black

Yellow

Purple

Blue

Orange

 

 

 

 

 

 

1.Plug a thermocouple into the input connector(s). (Make sure that the polarity is correct.)

2.Set the thermometer for the correct thermocouple type.

Displaying Temperatures

1.Press Cto select the correct temperature scale.

2.Hold or attach the thermocouple(s) to the measurement location.

The temperature reading appears in the primary display.

3.Model 52: Press Tto toggle between showing the T1, T2, and T1-T2 readings in the primary or secondary display.

The display shows 0L (overload) when the temperature being measured is outside the thermocouple’s valid range.

Model 52: If only thermocouple T2 is connected, the T2 reading appears in the primary display.

11

Image 15
Contents 51 & 52 Series Limited Warranty & Limitation of Liability Table of Contents 51 & 52 Series Contacting Fluke IntroductionWWarning Safety InformationIntroduction Getting Started International SymbolsComponents ComponentsDisplay Elements Display ElementsButtons Buttons51 & 52 Series Setup Options Using the ThermometerChanging Setup Options Entering or Exiting SetupChanging a Setup Option Type Color Measuring TemperaturesConnecting a Thermocouple Displaying TemperaturesUsing the Offset to Adjust for Probe Errors Holding the Displayed ReadingsViewing the MIN, MAX, and AVG Readings Specifications MaintenanceElectrical General 80 PK-1 Thermocouple supplied with ThermometerReplacement Parts and Accessories AccessoryPart Number 51 & 52 Series