Fluke 51 & 52 Series II user manual Changing a Setup Option

Page 14

51 & 52 Series II

Users Manual

Changing a Setup Option

1.Press Kor Nto scroll to the setup option you want to change.

2.Press Eto indicate that you want to change this setting.

3.Press Kor Nuntil the setting you want to use appears on the display.

4.Press Eto store the new setting in memory.

Notes

Setup is disabled in MIN MAX mode.

Offset:

The primary display shows the temperature plus the offset and the secondary display shows the offset. Remember to reset the offset to 0.0 when it is no longer needed. The offset automatically resets to 0.0 when you change the thermocouple type. Model 52: You can store individual offsets for T1 and T2.

Sleep mode:

The thermometer enters sleep mode if no button press occurs for 20 minutes. Pressing any button wakes the thermometer and returns it to its previous state. Sleep mode becomes enabled each time you turn on the thermometer and is automatically disabled in MIN MAX mode.

Line frequency:

For optimum rejection of line noise, the thermometer must be set for the local line frequency.

10

Image 14
Contents 51 & 52 Series Limited Warranty & Limitation of Liability Table of Contents 51 & 52 Series Introduction Contacting FlukeSafety Information WWarningIntroduction International Symbols Getting StartedComponents ComponentsDisplay Elements Display ElementsButtons Buttons51 & 52 Series Entering or Exiting Setup Using the ThermometerChanging Setup Options Setup OptionsChanging a Setup Option Displaying Temperatures Measuring TemperaturesConnecting a Thermocouple Type ColorViewing the MIN, MAX, and AVG Readings Using the Offset to Adjust for Probe ErrorsHolding the Displayed Readings Maintenance SpecificationsGeneral 80 PK-1 Thermocouple supplied with Thermometer ElectricalPart Number Replacement Parts and AccessoriesAccessory 51 & 52 Series