Anaheim H300 Battery Replacem ent, Each digit indicates the following status, RS232 Setting

Page 13

The 16 digit data str eam will be displayed in the following for mat:

D15 D14 D13 D12 D11 D10 D9 D8 D7 D6 D5 D4 D3 D2 D1D0

Each digit indicates the following status:

D15

Start Word

 

 

D14

4

 

 

 

D12 & D11

Indicator for Display

 

 

 

°C = 01

°F = 02

% RH = 04

D10

Polarity: 0 =

Positive / 1 =

Negative

 

Decimal Point (DP), position from right to left

D9

0

= No DP, 1 = 1 DP, 2 = 2 DP, 3 = 3 DP

 

Display reading, D1 = LSD, D8 = MSD

 

Ex ample: If the display reading is 1234,

D8 to D1

then D8 to D1 is: 00001234

 

 

 

 

 

D0

End Word

 

 

RS232 Setting

 

 

 

 

 

 

 

 

Baud Rate

9600

 

 

Parity

No Parity

 

 

Data bit #

8

Data bits

 

 

Stop bit

1

Stop bit

 

 

9. Battery Replacem ent

1.When the LCD display shows the “” low battery symbol, It is necessary to replace the battery. The specifications listed for this instrument do not apply when the low battery symbol is displayed.

2.Open the “ Battery Compartment” (4-7, Fig. 1) by first loosening the “ Battery Cover Screw s” (4-15, Fig. 1).

3. Replace the batteries with “ good” batteries noting the correct polarity . Replace

Battery Compartment” (4-7, Fig. 1) and “ Battery Cover Screws” (4-15, Fig. 1). Ensure the cov er is securely fastened.

12

Image 13
Contents Humidity/Temp Meter With DEW Point and Data Logger Page Auto Power Off Disable …………………………………… Advanced Measuring Procedures ………………………Send Data fr om RS232 Socket …………..………………… Applications Product Introduction FeaturesSafety Sym bols Safety Inform ationHumidity / Temper atur e SpecificationsDew Point Operation Instructions Unit DiagramHumidity & Tem perature Measurem ent Measuring ProcedureDEW Point Measurem ent Data HoldManual Data Logger set sample time to 0 seconds Change Unit of Measure from C to F Advanced Measuring ProceduresChange Data Logger Sampling Tim e Show Memory BalanceClear Mem ory Auto Power Off DisableSend Data fr om RS232 Socket RS232 PC Serial Interface Inform ationRS232 Setting Battery Replacem entEach digit indicates the following status Service Inform ation Lim ited Tw o-Year Warranty Anaheim Scientific