Anaheim H300 instruction manual Advanced Measuring Procedures, Change Unit of Measure from C to F

Page 10

5. Advanced Measuring Procedures

Note: Cancel any Hold or Record functions before trying to perform and advanced measuring procedures.

1.Press and hold the “ SET Button” (4-6, Fig. 1) for about 3 seconds until the LCD displays “ Code” . Once you release the button you should see 1000 in the main display. 1000 Is the password code to allow advanced measuring procedures. To leave the setting menu press the “ ESC Button” (4-3, Fig.1).

2.Press the “ Enter Button” (4-4, Fig. 1).

4.Now you are in the setting area to set or change the temperature unit of measure. Pressing the “ Set Button” (4-6, Fig. 1) will cycle through the following setting sections:

°F…… Change the Temp C or F (section 5-1)

SEC… Change the data logger sampling rate (section 5-2) Cnt….. To show the balance data in memory (section 5-3)

CLr …. Clear saved data in memory (data logger data) (section 5-4) OFF… Auto power ON/OFF management (section 5-5)

5-1 Change Unit of Measure from °C to °F

1.Press and hold the “ Set Button” (4-6, Fig. 1) for about 3 seconds until y ou see the w ord “ COde” in the bottom of the LCD.

2.Press the “ Enter Button” (4-4, Fig. 1). Now you may select betw een °C & °F by pressing the “Function Button” (4-5, Fig. 1). When a “ 0” is in the main LCD the unit w ill display temperature in °C. When a “ 1” is in the main LCD then the unit w ill display temperature in °F.

3.Once y ou hav e selected either a 0 or a 1, press the “ REC Enter” button and then press the “ Hold ESC” button to escape and go back to the normal testing mode.

5-2 Change Data Logger Sampling Tim e

1.Press and hold the “ Set Button” (4-6, Fig.1) for more than 3 seconds.

2.The main LCD should then display1000. Now press the “ Enter Button” (4-4-, Fig. 1)

3.Now press the “ Set Button” (4-6, Fig.1). You will see SEC in the bottom left of the LCD.

4.Use the “ Function Button” (4-6, Fig. 1) to cycle through the available seconds settings. Stop on either 0, 1, 2, 10, 30, 60, 600, 1800 or 3600. The value you select will tell the meter how often to log an event. Note that 0 SEC can only be used in the manual logging function.

5.Press the “ Enter Button” (4-4, Fig.1) to confirm your selection and save it to the meters memory.

6.Press the “ ESC Button” (4-3, Fig.1) to go to the normal measurement mode.

5-3 Show Memory Balance

1.Press and hold the “ Set Button” (4-6, Fig.1) for about 3 seconds.

2.The main LCD should then display1000. Now press the “ Enter Button” (4-4-, Fig. 1)

3.Press the “ Set Button” (4-6, Fig.1) twice and you will see the Cnt marker in the bottom of the LCD. The top number is the number of available data points.

4.Press the “ ESC Button” (4-3, Fig.1) to go to the normal measurement mode.

9

Image 10
Contents Humidity/Temp Meter With DEW Point and Data Logger Page Auto Power Off Disable …………………………………… Advanced Measuring Procedures ………………………Send Data fr om RS232 Socket …………..………………… Product Introduction Features ApplicationsSafety Inform ation Safety Sym bolsHumidity / Temper atur e SpecificationsDew Point Unit Diagram Operation InstructionsDEW Point Measurem ent Measuring ProcedureHumidity & Tem perature Measurem ent Data HoldManual Data Logger set sample time to 0 seconds Change Data Logger Sampling Tim e Advanced Measuring ProceduresChange Unit of Measure from C to F Show Memory BalanceAuto Power Off Disable Clear Mem oryRS232 PC Serial Interface Inform ation Send Data fr om RS232 SocketRS232 Setting Battery Replacem entEach digit indicates the following status Service Inform ation Lim ited Tw o-Year Warranty Anaheim Scientific