The ac source output is connected to the ac input of the equipment under test.

Set the output waveform to sine.

Set the rms voltage to Fixed mode.

Set the frequency to Fixed mode.

Set the voltage to 230 Vrms.

Set the frequency to 50 Hz.

Set the measurement window to Rectangular.

Enable the output of the ac power source/ analyzer.

Wait until the ac input to the equipment is in a steadystate condition.

Send the harmonic current array measurement command to return 50 harmonic current amplitudes.

Using the computer:

Read the harmonic current array into the computer.

Compare the measured harmonic current ampli- tudes of harmonic numbers 2 through 40 to the IEC 555-2 quasi-stationary standard limits.

19

Image 19
Contents Product Note Agilent 6800 Series AC Power Source/AnalyzerIntroduction Table of ContentsIntroduction to the Agilent 6800 AC Power Source/Analyzer Agilent 6800 Series Output Transients What is a Step Transient?What is a Pulse Transient? What is a List Transient?Model of the Transient System What Can Serve as the Source of the Trigger? What Actions Can be Triggered?How Can Triggers be Generated? Output Transient Trigger Model Measurement Trigger Model What is a FETCh Command? What is a MEASure Command?Agilent 6800 Series Measurements MEASure and FETCh Command Execution Diagram Application Transient Generation ApplicationsOverview of application Agilent 6800 series features usedAdvantages/benefits of the Agilent 6800 series solution Application Simulating AC Line Sub-Cycle DropoutsTiming Diagram of Application Page At 400 Hz Application Generating MIL-STD-704D WaveformsSet the List points as follows List Timing Diagram for Application Page Advantages/benefits of the Agilent 6800 series Application Performing IEC 555-2 MeasurementsUsing the computer Page Application Performing Inrush Current Measurements Agilent 6800 series setup Page How the 6800 series implements user-defined waveforms Application Generating User-Defined WaveformsTiming Diagram for Application #5 Page Page Advantages/benefits of the Agilent 6812A and 6813A solutions Agilent 6812A and 6813A features usedPage Our Promise