Agilent Technologies manual Agilent 6800 series setup

Page 22

Agilent 6800 series setup

Set the initial (immediate) voltage to 0 Vrms.

Set the triggered voltage level to 120 Vrms.

Set the frequency to 60 Hz.

Set the peak current limit to maximum.

Set the transient trigger source to BUS.

Set the trigger synchronization source to Phase.

Set the initial synchronization Phase to 40° (re-program to 50°, 60°, 70°, 80°, and 90°).

Set the acquisition trigger source to TTLTrg.

Set the data acquisition offset to -409 points.

Enable the output of the ac power source/ analyzer.

Initiate the transient trigger.

Initiate the acquisition trigger.

Send a Bus trigger.

Fetch the peak current measurement.

Upon receipt of the trigger, the ac power source/ analyzer output goes to 120 Vrms when the phase voltage waveform reaches 40°. The data buffer acquires 409 instantaneous current data points before the turn-on event and 3687 instantaneous current data points beginning at the onset of the turn-on event (for a total buffer of 4096 current data points) to generate the peak current measure- ment. To continue the peak inrush current charac- terization, change the synchronization phase (from 40° to 50°, 60°, 70°, 80°, and 90°) and repeat the same procedure.

Timing Diagram for Application 4

22

Image 22
Contents Agilent 6800 Series AC Power Source/Analyzer Product NoteTable of Contents IntroductionIntroduction to the Agilent 6800 AC Power Source/Analyzer What is a List Transient? What is a Step Transient?What is a Pulse Transient? Agilent 6800 Series Output TransientsModel of the Transient System What Can Serve as the Source of the Trigger? What Actions Can be Triggered?How Can Triggers be Generated? Output Transient Trigger Model Measurement Trigger Model What is a FETCh Command? What is a MEASure Command?Agilent 6800 Series Measurements MEASure and FETCh Command Execution Diagram Applications Application Transient GenerationApplication Simulating AC Line Sub-Cycle Dropouts Agilent 6800 series features usedAdvantages/benefits of the Agilent 6800 series solution Overview of applicationTiming Diagram of Application Page Application Generating MIL-STD-704D Waveforms At 400 HzSet the List points as follows List Timing Diagram for Application Page Application Performing IEC 555-2 Measurements Advantages/benefits of the Agilent 6800 seriesUsing the computer Page Application Performing Inrush Current Measurements Agilent 6800 series setup Page Application Generating User-Defined Waveforms How the 6800 series implements user-defined waveformsTiming Diagram for Application #5 Page Page Agilent 6812A and 6813A features used Advantages/benefits of the Agilent 6812A and 6813A solutionsPage Our Promise