Compur Gas Generator manual COCl2 and SO2 Generator see Picture

Page 13

4.2.3

COCl2 and SO2 Generator (see Picture 3)

1Unplug the external battery (COCl2 only).

2Loosen the 4 screws on the backside of the instrument and separate the two halves. (Use caution as the LED’s are attached to the top case)

3Remove the battery.

4Push back the 2 isolation tubes and unsolder the generator cell from the wires.

5Unscrew the clamp (KS) and push the generator cell backwards. Remove old generator cell.

6Insert the new generator cell and fasten the clamp.

7Solder the wires to the new generator cell (polarity of no importance) and push the isolation tubes back over the connections.

8Insert the battery.

9Close the housing and tighten the screws in the back. Reconnect the external battery pack.

Note:

)Whenever you have done maintenance work on a gas generator, test the function with a working gas detector. The detector should give an alarm within 10 seconds.

13

Image 13
Contents Page Nur bei COCl COCl 2 only Inhaltsverzeichnis / Table of Contents Testablauf Bedienungsanleitung FunktionsbeschreibungHinweis Batteriewechsel WartungBeachten Sie bitte dazu folgende Hinweise H2S und HCN Generator siehe Abbildung Austausch der GeneratorzelleCO, Cl2 und NO2 Generator siehe Abbildung COCl2 und SO2 Generator siehe Abbildung Operating Instructions Principle of Operation How To Run a Gas TestBattery Replacement MaintenancePlease observe the following notes H2S and HCN Generator see Picture Generator Cell ReplacementCO, Cl2 and NO2 Generator see Picture COCl2 and SO2 Generator see Picture Ersatzteile / Spare Parts Technische Daten / Technical DataGeneratorzellen / Generator cells Bestell-Nr Article-No Nur bei COCl2 / COCl2 onlyAbbildung / Drawing Abbildungen / Drawings H2S, HCN GeneratorCO, Cl2, NO2 Generator COCl2, SO2 Generator Notizen / Remarks