Delta LTL2000S manual Appendix B

Page 40

 

 

 

APPENDIX B

 

 

 

Status code interpretation

Rl Status code interpretation:

 

 

bit pattern

value

Description

Status Code:

00000001

1

Error in Rl Result

Status Code:

00000010

2

Stray Light Warning

Status Code:

00000100

4

Log full Warning

Status Code:

00001000

8

Not defined

Status Code:

00010000

16

Low Battery Warning

Status Code:

00100000

32

High Zero signal Warning

Status Code:

01000000

64

High Signal with Lamp on Error

Status Code:

10000000

128

High Signal with Lamp off Error

Qd Status code interpretation:

 

 

bit pattern

value

Description

Status Code:

00000001

1

Error in Qd Result

Status Code:

00000010

2

Stray Light Warning

Status Code:

00000100

4

Not defined

Status Code:

00001000

8

Not defined

Status Code:

00010000

16

Low Battery Warning

Status Code:

00100000

32

Not defined

Status Code:

01000000

64

High Signal with Lamp on Error

Status Code:

10000000

128

High Signal with Lamp off Error

E.g. Status code 18 is composed of Low Battery Warning (16) + Stray Light Warning (2)

DELTA

LTL2000S(Q) Retroreflectometer

36

Image 40
Contents LTL2000S / SQ Retroreflectometer Disclaimer Index Specification Introduction LTL2000SLTL2000SQ features Getting started Buttons MeasuringHome Wet night measurement considerationsScroll Keys Calibration LTL2000S Rl CalibrationDisplay shows Step Calibrate Calibration LTL2000SQQd Calibration How to place the calibration normals Rl CalibrationRemember Rl Measurement LTL2000SQ RL optical system layoutDelta A road marking 500 400 300 200 100 Up to 5 mm profile height Up to 20 cm profile gap Qd Measurement 516 517 512 Instrument function Factory calibrationDelta Keyboard functions Keyboard LayoutKey Label Key Function Clear log data Measurement idQd Mode Rl ModeMean Calculation Menu systemQd Signal Qd Stray L VBat QdDisable Id Default SettingsOff Timer Print LogGeneral care BatteryProtection window Fuses LampCalibration unit Rl calibration unit CeramicsLight trap Rl calibration unit Light trapHow to replacing the printer paper PrinterRS-232C specification LTL DTE Data protocol Command format Command set Following commands have been definedTEST2 Command Parameter Response Appendix B Rl Calibration Printout Result PrintoutQd Calibration Printout Log Printout Status Printout Qd Calibration data Qd measurement dataGeneral Characteristics Rl General Characteristics QdElectrical Characteristics