Page 4

Mounting continued

Mounting on a Center Pole

1Power Supply Module & Battery Module Mounting

1A Attach the Power Supply Module to the mounting bracket using the included hardware.* Make sure the Power Supply Module is oriented as shown in the diagram, with its outlets facing up.

Attach included hardware to the bottom of the mounting bracket.** Fully tighten the screws. (NOTE: These screws are designed to protrude downward even when fully tightened in order to engage the cart’s legs in step 1B.)

*Four phillips-head screws (3/8 in. long, 10-32 size) and four star washers (#10).

**Two phillips-head screws (1 in. long, 10-32 size). Two fender washers (#10).

1B Using one or more assistants, lift the Power Supply Module/mounting bracket and place it around the cart's center pole as shown. (NOTE: Since the mounting bracket is designed to accommodate a wide variety of center pole sizes, attach the screw clamps as described in step 1C.) When lowering the mounting bracket into place, make sure that the screws on the bottom of the bracket engage the cart's legs in order to prevent the retrofit kit from rotating around the cart's center pole. To minimize rotation, make sure the side of the mounting bracket is perpendicular (90°) in relation to one of the cart's legs, as shown in the “Top View” diagram. Lift and readjust if needed.

1C The mounting bracket is designed to accommodate a wide variety of center pole sizes. To secure bracket to center pole, wrap the two included screw clamps around the cart's center pole and through the raised slots on the mounting bracket. Use a screwdriver to fully tighten clamps.

1D Using one or more assistants, lift the Battery Supply Module and attach it to the Power Supply Module/mounting bracket as shown using the mounting flanges and included hardware.*

*Four hex bolts (1-5/8 in. long, 5/16-18 size), four hex nuts (5/16-18), four star washers (5/16) and four flat washers.

1A

 

Top View

 

 

CENTER

LEG

90°

 

 

 

POLE

 

 

 

 

1B

MOUNTING BRACKET

 

 

 

1C

1D

4

Image 4
Contents Medical-Grade Mobile Power Retrofit Kit Power Supply Module Connection Warnings Power Supply Module & Battery Module Location WarningsEquipment Connection Warnings Battery Module Warnings Labeling SymbolsLEG Remote User Interface RUI Mounting Mounting Plug In the Power Supply Module Connect the Battery Module to the Power Supply ModuleCharge the Battery Module Plug Your AC Equipment into Power Supply Module’s Outlet Turn the Power Supply Module’s Outlet onOPTIONAL-Install PowerAlert Data-Saving Software Mobile Operation Unplugged / Battery Discharging Stationary Operation Plugged In / Battery ChargingRemote User Interface RUI Power Supply Module Battery Module Power button after reconnection. See Connection / Start-Up Service StorageLimited Warranty Page 200702102