Tripp Lite HCRK-INT owner manual Remote User Interface RUI Mounting

Page 5

Mounting continued

2 Remote User Interface

(RUI) Mounting

Mount the RUI to the top edge of the mounting bracket with included screws.

Or, if desired, use the included double-sided tape to secure the RUI to any flat surface on

the cart. Read caution on page 6 concerning 2 double-sided tape.

Mounting on a Bottom Shelf (Replacing Existing Shelf)

Note: The dimensions of the included replacement shelf are 22 inches x 17.26 inches [55.88 cm. x 43.84 cm.]

1Power Supply Module & Battery Module Mounting

1A Lay cart on its side and remove existing shelf from bottom of cart.*

*The existing shelf is not required for mounting or operation of Tripp Lite's Retrofit Kit. However, Tripp Lite's replacement shelf includes holes that will accept the blocks from the existing shelf (if desired).

1B Attach Power Supply Module to replacement mounting shelf with included hardware.* Make sure that the Power Supply Module's outlets are facing opposite of where you will mount the Battery Module in step 1D. NOTE: the mounting tray features identical mounting holes on each side, allowing Power Supply Module and Battery Module to be mounted on either side of the cart.

*Power Supply Hardware: four phillips-head screws (3/8 in. long, 10-32 size) and four star washers (#10).

1C Attach replacement mounting shelf to bottom of cart with included hardware.* To securely engage the cart's threaded holes, the included bolts must be screwed from the outside in (as shown). Access the cart's threaded holes by reaching inside the hollow crossbar that holds the cart's wheels.

*Mounting Tray Hardware: ten hex bolts (3/8 in. long 1/4-20 size) and 10 lock washers.

1D Stand cart upright. Attach Battery Module to replacement mounting shelf with included hardware.*

*Battery Module Hardware: four hex bolts (1-5/8 in. long 5/16-18 size), four hex nuts (5/16-18), four star washers (5/16) and four flat washers.

1A

1B

1C

1D

5

Image 5
Contents Medical-Grade Mobile Power Retrofit Kit Equipment Connection Warnings Power Supply Module & Battery Module Location WarningsPower Supply Module Connection Warnings Labeling Symbols Battery Module WarningsLEG Remote User Interface RUI Mounting Mounting Charge the Battery Module Connect the Battery Module to the Power Supply ModulePlug In the Power Supply Module OPTIONAL-Install PowerAlert Data-Saving Software Turn the Power Supply Module’s Outlet onPlug Your AC Equipment into Power Supply Module’s Outlet Stationary Operation Plugged In / Battery Charging Mobile Operation Unplugged / Battery DischargingRemote User Interface RUI Power Supply Module Battery Module Power button after reconnection. See Connection / Start-Up Storage ServiceLimited Warranty Page 200702102