8.3 Dynamic Characteristics
8.3.1 Output Voltage Deviation
Output Voltage Deviation is defined as the response of the converter to a sudden step change in the output load current. The output voltage deviation is characterized by two parameters: Maximum Output Voltage Deviation and Response Time (please refer to the data sheet for the detailed specification). The value of dynamic resistance for a defined step current is defined as:
Rdynamic | = | Vout | |
0.5 * Imax | |||
|
|
Test
1)Turn on the fan.
2)Adjust the input voltage to the desired operating point.
3)Set the electronic or resistive load at 50% of maximum load.
4)Change channel 1 to scope probe and measure across the Rdynamic .
5)Set the switch SW1 to the “OFF” position to enable the converter.
6)Set channel 2 on the oscilloscope to be AC coupled and to 50mV/Div and for 50uS/Div. Set the trigger to auto and adjust the trigger point at a negative going pulse for step load change from 50% to 100% of Io or adjust the trigger point at positive going pulse for step load change from 100% to 50% (Please refer to data sheet)
7)Measure the Peak deviation and capture the waveform as required.
8.3.2 Turn-On Transient Time
1)Turn on the module by using the External switch to control input voltage.
2)Turn on module by using the Enable on/off.
Note: There is a difference in performance in each mode - please refer to the data sheet for the detailed specification.
11