Hardware Reference - Flextight 343 | 17 | |
The Light Tube | ||
|
Replacing the Light Tube
If your scans are suddenly coming out black (or all white if you are scanning negatives), then the light tube may be defective or misaligned.
Follow procedure below to replace a tube:
1.Get the appropriate replacement light tube (ask your Flextight dealer or see specification in “Technical Specifications” on page 20).
2.Turn off and unplug the scanner from the primary power supply.
3.IMPORTANT: Let the light tube cool down for app. 10 minutes before proceeding.
4.Lay the scanner down on its back, with the bottom facing you.
5.Use a 2.5 mm allen key to remove the screws located at each of the four corners of the bottom panel (see Figure 6), then remove the bottom panel.
English
1
4
3
2