Omega Engineering DAQ-12 user manual Start of Conversion Register, 3 A/D Converter Data Register

Page 27

VALID - when set, logic 1, indicates at least one data sample was lost because it was read by the personal computer before the next sample was converted. Data was lost because the sampling rate was too fast for the computer to acquire the data. VALID is reset by writing to the start conversion register.

CHSL2, CHSL1, CHSL0 - select the multiplexer channel for the analog input signal. (* denotes only available in single ended input mode).

CHSL3

CHSL2

CHSL1

CHSL0

MUX channel

0

0

0

0

channel 0

0

0

0

1

channel 1

0

0

1

0

channel 2

0

0

1

1

channel 3

0

1

0

0

channel 4

0

1

0

1

channel 5

0

1

1

0

channel 6

0

1

1

1

channel 7

1

0

0

0

channel 8*

1

0

0

1

channel 9*

1

0

1

0

channel 10*

1

0

1

1

channel 11*

1

1

0

0

channel 12*

1

1

0

1

channel 13*

1

1

1

0

channel 14*

1

1

1

1

channel 15*

4.1.2Start of Conversion Register

The start of conversion register is 16-bit write only and performs two functions:

1.When configured for internal triggering, writing a "0" to this register generates the software trigger, starting the data conversion process.

2.Writing a "0" to this register at any time resets the VALID bit in the control word register. This allows the VALID bit to be reset at any time during the conversion

process or before the event of an external trigger.

4.1.3A/D Converter Data Register

An input to this register returns the last digital value converted by the A/D converter. This register is 16-bit read only.

DAQ-12 Users Manual

27

Image 27
Contents 15,1*,1 WARRANTY/DISCLAIMER Standards to which Conformity is Declared Manufacturers NameManufacturer’s Address Application of Council DirectiveServicing North America For immediate technical or application assistanceServicing Europe Page Table of Contents Introduction List of Figures and Tables Installation DAQ-12 Specifications Analog to Digital Converter Jumper J1 ConfigurationDisable Pre-Scaler Enable Pre-Scaler Factory default Recommended Input Ranges and Gain Settings Code Voltage Unipolar Bipolar Code Digital to Analog ConvertersGain = Channel BipolarBase Address Digital Input/OutputClock Selection Internal Clock External Clock Start of Conversion Trigger Selection10. Jumpers J8 and J9 Configuration Dack Direct Memory AccessExternal Interrupt InterruptsExternal Connections Page Control Word Register Register DescriptionINT2 INT1 INT0 Description 3 A/D Converter Data Register Start of Conversion RegisterClock Rate Register low word Multi-Function Timer Register4 D/A Converter 0 Register 5 D/A Converter 1 RegisterProgramming the 8254 Counter/Timer Counter 1 Clock rate register high word Version January 28