Agilent Technologies 664xA, 665xA, 667xA, 669xA, 668xA manual Connecting the Sense Leads

Page 71

Connecting the Sense Leads

You must connect the positive side of the load to the +S analog connector pin and the negative side of the load to the -S analog connector pin (see Figure 4-1). Connect the sense leads carefully so that they do not become open-circuited. If sense leads are left open during operation, the supply will regulate at the output terminals instead of at the load. Remember to bundle or tie wrap the load leads to minimize inductance and reduce noise pickup.

CV Regulation

The voltage load regulation specification in Table 1-3a applies at the output terminals of the power supply. When remote sensing, this specification must be compensated. Add an increment to the voltage load regulation specification as specified by “∆mV” in the equation given under Load regulation in Table 1-3b.

OVP Considerations

The OVP circuit senses the voltage near the output terminals and not at the sense terminals. Depending on the voltage drop between the output terminals and the load, the voltage sensed by the OVP circuit can be significantly higher than actually being regulated at the load. You must program the OVP trip high enough to compensate for the expected higher voltage at the output terminals.

Output Rating

The rated output voltage and current specification in Table 1-3a applies at the output terminals of the power supply. With remote sensing, any voltage dropped in the load leads causes the supply to increase the voltage at the output terminals so it can maintain the proper voltage at the load. When you attempt to operate at the full-rated output at the load, this forces the supply voltage at the output terminals to exceed the supply's rated output. This will not damage the supply, but may trip the OVP (overvoltage protection) circuit, which senses the voltage at the output bus bars. When operated beyond its rated output, the supply's performance specifications are not guaranteed, although typical performance may be good. If the excessive demand on the supply forces it to lose regulation, the Unr annunciator will indicate that the output is unregulated.

Output Noise

Any noise picked up on the sense leads also appears at the output of the power supply and may adversely affect the load voltage regulation. Be sure to twist the sense leads to minimize external noise pickup and route them parallel and close to the load leads. In noisy environments, it may be necessary to shield the sense leads. Ground the shield only at the power supply. Do not use the shield as one of the sense conductors.

Note

The signal ground binding post on the rear panel is a convenient place to ground the sense shield.

Stability

Using remote sensing under unusual combinations of load-lead lengths and large load capacitances may cause your application to form a low-pass filter that becomes part of the voltage feedback loop. The extra phase shift created by this filter can degrade the supply's stability and result in poor transient response. In severe cases, this may cause output oscillations. To minimize this possibility, keep the load leads as short as possible and tie wrap them together.

In most cases, following the above guidelines will prevent problems associated with load lead inductance. However, if a large bypass capacitor is required at the load and load-lead length cannot be reduced, then a sense-lead bypass network may be needed to ensure stability (see Figure 4-4b). The voltage rating of the 33 ∝F capacitors should be about 50% greater than the anticipated load-lead drop. Addition of the 20-Ω resistors will cause a slight voltage rise at the remote sensing points. For utmost voltage programming accuracy, the supply should be recalibrated with the DVM at the remote sensing points (see “Appendix A - Calibration”).

Note

If you need help in solving a stability problem with any Series 667xA power supply contact an Agilent

 

Service Engineer through your local Agilent Sales and Support Offices.

User Connections 71

Image 71
Contents Agilent Part No Microfiche Part No January Operating Guide Gpib DC Power SuppliesExclusive Remedies WarrantyLimitation of Warranty CertificationGeneral Safety SummaryHerstellerbescheinigung Safety Symbol DefinitionsSymbol Description Printing HistoryManufacturer’s Name and Address Declaration of Conformity6x4yA6x5yAE435xA.b.11.24doc.doc 6x4yA6x5yAE435xA.a.11.24doc.doc Table of Contents Introduction Getting Acquainted Programming the Output Connecting Series 667xA Power Supplies to the LoadController Connections Calibration Over the Gpib 100 Introduction Equipment Required General ProcedureOption 601 Installation 135 Option 602 Installation 136 Front Panel CalibrationTopic Location General InformationIntroduction Quick Document OrientationOptions Safety ConsiderationsInstrument Identification Support rails E3663AC are required AccessoriesFamily Power Agilent Models DescriptionAnalog Programming Front Panel ProgrammingRemote Programming General Output CharacteristicSpecifications and Supplemental Characteristics Series Specifications CharacteristicsCurrent@ 55C Output Ratings VoltageCurrent@ 40C Current@ 50COutput Programming Range maximum Average Resolution VoltageTemperature Coefficients change per C Voltage Maximum Input PowerMaximum Reverse Bias Current Auto-Parallel ConfigurationAnalog Programming IP & VP Input Signal Maximum AC Line Current Ratings Vac nominalGpib Interface Capabilities Digital Port CharacteristicsSafety Compliance Complies with Dimensions WidthOutput Impedance Curves Typical General Information 50 a 25 a 15 a 51.188 a 25.594 a 15.356 a 214 a 095 a Rms 15 AM fuse Digital Port Characteristics Output Impedance Curves Typical General Information ±Current Current@ 0 to 55CVoltage 04% + Current0 . l % + Analog Programming ±IP Current Monitor +IM Typical Resolution VoltageMaximum Input VA and Power ±Current ReadbackFull-load programming speed up/down time time for Analog Programming IP & VP Input Signal VP Input ImpedanceIP to -IP Differential Input Signal Current Monitor Output +IM Output SignalOutput Characteristic Curve General Information Weight NetOutput Impedance Curves Typical General Information Output Ratings Voltage Current Typical Common Mode Noise Current Rms Peak-to-peak VP Input Signal+ IP Input Signal Current Monitor IM Output SignalRange Rms line current Maximum Reverse Voltage Current Sink CapabilityMaximum AC Line Current Ratings Range Rms line current Line fuseHeight Maximum memory write cyclesDesigned to comply with Milliohms Output Impedance Curves Typical440 a 220 a 110 a Derated linearly 1%/C from 40 C to Temperature Coefficients change per C Max Power 6.67KW VoltageOvervoltage Protection OVP Typical Resolution Voltage Drift Temperature Stability9000 VA Output Characteristic Curve Vout Agilent 6692A Agilent 6690AAgilent 6691A Operator Replaceable Parts List Description Agilent Part No Supplemental Gpib Characteristics for All Models ParameterOperator Replaceable Parts List Description Agilent Part No Items Supplied InstallationInspection Damage Packaging MaterialBench Operation Location and TemperatureCleaning Temperature PerformanceInstalling the Series 667xA Power Cord Input Power SourceInstalling the Series 664xA and 665xA Power Cord Connecting the Series 667xA Power Cord Series 668xA/669xA Overall Wiring Diagram Installing the Series 668xA Power CordConnecting the Series 668xA Power Cord Installing the Series 669xA Power CordConnecting the Series 669xA Power Cord Page Turn-On Checkout Turn-On CheckoutIntroduction Preliminary Checkout All Models Shifted Keys Power-On Checkout All ModelsUsing the Keypad All Models Output Checkout All ModelsPress Prot Clear Checking the Voltage FunctionOpen or Connected to a Voltmeter Checking the Current Function Line Fuse Checking the Save and Recall Functions All ModelsDetermining the Gpib Address All Models Case of TroubleSeries 667xA Supplies Series 664xA and 665xA SuppliesSeries 668xA Supplies Error Messages All ModelsSelftest Errors Power-On Error MessagesError Display Failed Test Checksum ErrorsRuntime Error Messages Power-On Selftest Errors Display Failed TestRear Panel Connections All Models Load Wire Selection All ModelsUser Connections Pin No Fault/Inhibit Digital I/O Analog Connector All ModelsDigital Connector All Models 6651A 6652A 6653A 6654A Output IsolationLoad Considerations Capacitive Loads6641A 6642A 6643A 6644A 6645A 6651A 6652A 6653A 6654A 6655A Battery ChargingLocal Voltage Sensing Inductive LoadsRemote Voltage Sensing Stability Connecting One Supply to the LoadConnecting Supplies in Auto-Parallel Insert Protection Diodes Enable OCP on the MasterProgram Slave 2 OVP to the Maximum Level External Voltage Control Connecting Supplies in SeriesWiring Considerations Programming Connecting Series 667xA Power Supplies to the Load4a. Series 667xA Rear Panel Output Connections 6671A 6672A 6673A 6674A 6675A Connecting the Sense Leads Connecting One Power Supply To Multiple Loads Connecting One Power Supply to a Single LoadConnecting Supplies in Auto-Parallel 4f. Series 667xA Series Connection Remote Sensing Optional 4g. Series 667xA Analog Programming Connections Wiring Considerations -4gConnecting Series 668xA and 669xA Power Supplies to the Load Local Voltage Sensing Remote Sense Points Load LeadsInstructions supplied with the kit Connecting Supplies in Auto-Parallel Connecting Supplies in Series 5g. Series 668xA and 669xA Analog Programming Connections Linked Connections Controller ConnectionsStand-Alone Connections Controller Connections Front Panel Operation Front Panel OperationGetting Acquainted Dis DisplayStatus Annunciators UnrFunction Keys Front Panel Controls and IndicatorsOutput Rotary Controls Voltage System KeysEntry Keys Thru Press to select numerical values Line Switch On / OffProgramming the Output Establishing Initial ConditionsProgramming Overvoltage Protection Setting the OVP LevelProgramming Voltage Clearing The OVP Condition Programming CurrentChecking OVP Operation Checking OCP Operation CV Mode vs. CC ModeSetting The OCP Protection Programming Overcurrent ProtectionTurn-On Conditions Unregulated OperationSaving and Recalling Operating States Action Display Shows Setting the Gpib AddressTypes of Power Supply Gpib Addresses Changing the Power Supply Gpib AddressPage Parameters Calibrated General ProcedureCalibration Equipment RequiredFront Panel Calibration Figure A-1. Calibration Test Setup Series 668xA/669xA SetupEntering Current Calibration Values Enabling the Calibration Mode PASWDlEntering Voltage Calibration Values Calibrating the OVP Trip PointTable A-3. Gpib Calibration Error Messages Meaning Recovering From Calibration ProblemsCalibration Error Messages Calibration Language Dictionary Front Panel Corresponding Scpi CommandCalibration Over the Gpib Calcurrmon Series 668xA/669xA only Command SyntaxCommand Syntax CALibrateSAVE Parameters None Examples Agilent Basic Calibration Program CalvoltFigure A-2. Agilent Basic Calibration Program Steps 640 Through 670 not Used on 664x, 665x 570 ! Line 590 Password Must be Edited for Model Other thanCurrent Monitoring Resistor Operation VerificationTest Equipment Required List of EquipmentFigure B-1. Verification Test Setup Performing the Tests Sufficient size to carry the maximum rated current Current Programming and Readback AccuracyModel Agilent 6643A Voltage Programming and Readback Model Agilent 6641A Voltage Programming and ReadbackCurrent Programming and Readback Model Agilent 6642A Voltage Programming and ReadbackModel Agilent 6645A Voltage Programming and Readback Model Agilent 6654A Voltage Programming and Readback Model Agilent 6651A Voltage Programming and ReadbackModel Agilent 6652A Voltage Programming and Readback Model Agilent 6653A Voltage Programming and ReadbackModel Agilent 6655A Voltage Programming and Readback Model Agilent 6674A Voltage Programming and Readback Model Agilent 6671A Voltage Programming and ReadbackModel Agilent 6672A Voltage Programming and Readback Model Agilent 6673A Voltage Programming and ReadbackModel Agilent 6675A Voltage Programming and Readback Model Agilent 6683A Voltage Programming and Readback Model Agilent 6680A Voltage Programming and ReadbackModel Agilent 6681A Voltage Programming and Readback Model Agilent 6682A Voltage Programming and ReadbackModel Agilent 6684A Voltage Programming and Readback Model Agilent 6692A Voltage Programming and Readback Model Agilent 6690A Voltage Programming and ReadbackModel Agilent 6691A Voltage Programming and Readback Page Line Voltage Conversion Series 664xA and 665xA Power SuppliesLine Voltage Conversion Figure C-2. Series 665xA Line Select Jumpers Series 667xA Power SuppliesSeries 668xA/669xA Power Supplies Figure C-4. Removing the Series 668xA/669xA Inner Cover Digital Port Functions Digital ConnectorFault/Inhibit Operation Figure D-2. Example of Inhibit Input Figure D-3. Examples of FLT Outputs Common pin Changing the Port ConfigurationDigital I/O Operation IN/OUT 2 pinCommon Relay Link OperationPage Current Loop Compensation Series 668xA Only Current Loop Compensation Series 668xA OnlyFunction of Loop Compensation Current Loop Compensation Series 668xA Only Current Loop Compensation Series 668xA Only Figure E-1. CC Loop Compensation Curves For Model 6684A Setting the Loop Compensation SwitchAutoparallel Procedure Figure F-1 Master/Slave Current Division Using Agilent 668xA Series Power Supplies in AutoparallelOutput Bus Bar Options Option 601 InstallationOutput Bus Bar Options Option 602 InstallationBus Bar Spacer, 5040-1699 Minus Bus Bar Plus Bus Bar Customer bus railsIndex IndexGpib 6665xA, 24 667xA, 29 668xA, 34 669xA, 39 output isolation Index Japan United States Latin AmericaCanada Australia/New Zealand Europe Asia PacificManual Updates
Related manuals
Manual 70 pages 51.66 Kb Manual 67 pages 46.05 Kb

668xA, 669xA, 667xA, 664xA, 665xA specifications

Agilent Technologies has long been a pioneer in the production of high-performance electronic test and measurement instruments, particularly in the field of power sources. Among its notable offerings are the Agilent 667xA, 669xA, 665xA, 664xA, and 668xA series of power supplies. These instruments are designed to provide stable, reliable power for a variety of applications, including electronic testing, industrial processes, and research laboratories.

The Agilent 667xA series is characterized by its programmability and advanced measurement functions. These power supplies support a wide range of output voltages and currents, allowing for flexible configurations that cater to different testing needs. The built-in measurement capabilities enable users to monitor the voltage, current, and power with high precision, which is essential for ensuring optimal performance in electronic applications.

The Agilent 669xA series stands out with its high-power outputs, making it suitable for demanding applications. These power supplies deliver high voltage and current levels, making them ideal for testing high-performance devices, such as power amplifiers and motor drives. Additionally, the 669xA series includes features such as overvoltage protection and complex output sequencing to enhance the safety and reliability of the testing process.

The Agilent 665xA and 664xA series focus on delivering high accuracy and excellent regulation. These models are particularly known for their low noise operation, which is critical for sensitive applications where precision is paramount. The integrated programming capabilities allow users to automate testing sequences, thus improving efficiency in research and development settings.

The 668xA series features advanced digital signal processing that enhances the precision and stability of the output. Users benefit from features like remote sensing and monitoring, allowing feedback adjustments that maintain output accuracy despite cable losses. Furthermore, the 668xA models can integrate seamlessly with various test environments thanks to their LAN, GPIB, and USB connectivity options.

Overall, the Agilent 667xA, 669xA, 665xA, 664xA, and 668xA power supplies provide a comprehensive range of solutions for diverse electronic testing needs. With their advanced features, superb measurement capabilities, and robust performance, these instruments empower engineers and researchers to conduct their work with confidence, precision, and efficiency.