Anritsu MS2038C High Port Power, Measurement Uncertainties, Frequency Range GHz Directivity dB

Page 5

High Port Power

OSLxx50 Calibration Components (N-Connectors)

Corrected System Performance and Uncertainties: MS202x/3xC Model with 12-term SOLT calibration including isolation using either OSLN50 & OSLNF50 or OSLK50 & OSLKF50 Calibration Kits

Precision calibration standards come in a convenient configuration for field work.

Frequency Range (GHz)

Directivity (dB)

 

 

≤ 5

> 42

 

 

≤ 15

> 36

 

 

≤ 20*

> 32

 

 

* N Connector guaranteed to 18 GHz, typical > 18 GHz

Frequency Range (GHz)

Typical High Port Power (dBm)

 

 

≤ 3

+3

 

 

≤ 6

–3

 

 

≤ 20

–3

 

 

Measurement Uncertainties

The following graphs provide measurement uncertainty at 23 ºC ± 5 ºC for the above indicated connector type and calibration. Errors are worse-case contributions of residual directivity, source match, frequency response, network analyzer dynamic range, and connector repeatability. For two-port measurements, transmission tracking, crosstalk, and physical load match termination were added. Isolation calibration and an IF Bandwidth of 10 Hz is used.

S11 Magnitude

10

Uncertainty 0.5 MHz to 20 MHz

Uncertainty 20 MHz to 3 GHz

Uncertainty 3 GHz to 6 GHz

Uncertainty 6 GHz to 20 GHz

1

0.1

 

 

 

 

 

 

 

 

-40

-35

-30

-25

-20

-15

-10

-5

0

S11 Phase

100

Uncertainty 0.5 MHz to 20 MHz

Uncertainty 20 MHz to 3 GHz

Uncertainty 3 GHz to 6 GHz

Uncertainty 6 GHz to 20 GHz

S21 Magnitude

10

Uncertainty 0.5 MHz to 20 MHz

Uncertainty 20 MHz to 3 GHz

Uncertainty 3 GHz to 6 GHz

Uncertainty 6 GHz to 20 GHz

1

0.1

0.01

-80

-70

-60

-50

-40

-30

-20

-10

0

S21 Phase

100

Uncertainty 0.5 MHz to 20 MHz

Uncertainty 20 MHz to 3 GHz

Uncertainty 3 GHz to 6 GHz

Uncertainty 6 GHz to 20 GHz

10

10

1

1

 

 

 

 

 

 

 

 

 

0.1

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

-40

-35

-30

-25

-20

-15

-10

-5

0

-80

-70

-60

-50

-40

-30

-20

-10

0

Image 5
Contents High Performance Handheld S-Parameters IntroductionVNA Master + Spectrum Analyzer Performance and Functional Highlights VNA MasterVNA Master Functional Specifications 20 dB Frequency Range GHz Typical High Port Power dBm High Port PowerMeasurement Uncertainties Frequency Range GHz Directivity dBLow Port Power High Port Power Frequency Range GHz Typical Low Port Power dBm Bandwidth FrequencySpectral Purity Amplitude RangesSecond Harmonic Distortion P1dBVNA Performance Capabilities MS202x/3xC + Spectrum Analyzer Performance Capabilities Secure Data Operation Option Power Monitor Option 0005 Requires external detectorPower Monitor Detectors* Ordered separately Bias Tee Option Internal Resolution External Voltage/CurrentVector Voltmeter Option Internal Voltage/CurrentChannel Scanner Option 0027 Models MS2036/38C only Distance Domain Option 0501 included in Time Domain Option Balanced/Differential S-Parameters, 1-port OptionVNA Master General Specifications MS202x/3xC Input and Output Ports SpecificationsGeneral Specifications Master Software Tools and Remote Programming Feature BenefitAncillary Equipment MS202x/3xC Standard AccessoriesOrdering Information Optional Accessories Manuals Waveguide Calibration Components and WG/ Coaxial AdaptersRelated Literature, Application Notes Part NumberPage Canada Anritsu CompanyBrazil Mexico Anritsu Company, S.A. de C.V