Measurement Options Specifications
Time Domain (Option 0002) (includes Distance Domain Option 0501)
The VNA Master can also display the
Distance Domain
(0.5 x c x Vp) / ∆F; (c is speed of light = 3E8 m/s, ∆F is F2 – F1 in Hz) | |
|
|
(c x Vp) / ∆F; (c is speed of light = 3E8 m/s, ∆F is F2 – F1 in Hz) | |
|
|
Horizontal Range (meters): | 0 to (data points – 1) x Fault Resolution to a maximum of 3000 m (9843 ft.) |
|
|
Windowing | Rectangular, Nominal Side Lobe (NSL), Low Side Lobe (LSL), and Minimum Side Lobe (MSL) |
|
|
Power Monitor (Option 0005) Requires external detector
Transmitter measurements in the field are possible when using this VNA Master software mode with a separately purchased Anritsu 560 series detector. A variety of detectors are available to 50 GHz, but the popular
+20 dBm with better than 0.5 dB flatness to 18 GHz. After zeroing the detector to ensure accuracy at low power levels, the software offers intuitive operation for absolute and relative readouts in dBm or Watts.
Display Range | ||
|
| |
Measurement Range | ||
|
| |
Offset Range | 0 to +60 dB | |
|
| |
Resolution | 0.1 dB, 0.1 xW (x = n, µ, m based on detector power) | |
|
| |
Accuracy | ±1 dB maximum for | |
| ||
|
|
Power Monitor Detectors* (Ordered separately):
Part Numbers | |||
|
|
| |
Frequency Range | 0.01 to 20 GHz | 0.01 to 20 GHz | |
|
|
| |
Impedance | 50 Ω | 50 Ω | |
|
|
| |
Power Range | |||
|
|
| |
| 15 dB, < 0.04 GHz | 15 dB, < 0.04 GHz | |
Return Loss | 22 dB, < 8 GHz | 22 dB, < 8 GHz | |
17 dB, < 18 GHz | 17 dB, < 18 GHz | ||
| |||
| 14 dB, < 20 GHz | 14 dB, < 20 GHz | |
|
|
| |
Input Connector | N(m) | WSMA(m) | |
|
|
| |
Frequency Response | ±0.5 dB, < 18 GHz | ±0.5 dB, < 18 GHz | |
±1.25 dB, < 20 GHz | ±1.25 dB, < 20 GHz | ||
| |||
|
|
|
*See www.us.anritsu.com for additional detectors
Secure Data Operation (Option 0007)
For highly secure data handling requirements, this software option prevents the storing of measurement setup or data information onto any internal file storage location. Instead, setup and measurement information is stored ONLY to the external USB memory location. A simple factory preset prepares the VNA Master for transportation while the USB memory remains behind in the secure environment. The VNA Master cannot be switched between secure and