J E O L J S M - 6 0 6 0 L V S C A N N I N G E L E C T R O N M I C R O S C O P E

4

Instrument Operation

his section will walk you through the basic operation of the scanning electron Tmicroscope. You will follow a series of tasks that includes

Powering up the instrument,

Loading your sample,

Evacuating the chamber,

Turning on the electron gun, and

Gazing at some wonderful high magnification images.

Instrument Startup

1.If the instrument is already turned on, skip to the next section on Sample Loading. If the instrument is off, continue to Step 2.

2.Turn on the water chiller. The on/off switch is on the front panel. The temperature should read between 65 and 69 degrees F.

3.Turn on the SEM by turning and releasing the key on the front panel of the instrument (it’s just like starting a car).

4.Turn on the Dell computer on the left and hit the ENTER key at the log on screen (no password necessary).

Sample Loading

1.If the chamber is evacuated (EVAC button on the front of the instrument is lit green), press and hold the VENT button until it lights up yellow. This will vent the chamber.

2.Wait for the VENT button to stop flashing (about 40 seconds).

3.Open the chamber by grabbing the door assembly at the side depressions and carefully sliding the door assembly outwards. You should be able to see the brass, circular platform of the sample stage when you slide open the door.

4.Slide your sample holder with sample attached onto the copper-colored dovetail on the stage.

5.CAUTION: Adjust the sample height so that the top of the sample is well below the position of the interior chamber components (detectors, lenses, etc.). Use the z-axis control (the adjustment on the door assembly with the knob that is pointing straight up and the micrometer with horizontal divisions) for height adjustments. Steer clear of anything inside the chamber.

Bumping interior components can cause thousands of dollars in damage to the instrument. Know your sample height, and know approximately how much clearance you have around your sample, particularly along the z-axis.

7

Image 8
Contents Operating Instructions Table of Contents Safety IntroductionBackground References Samples Sample HoldersSample Preparation Conductive Materials Sample Preparation Non-conductive MaterialsPage Sample Loading Instrument OperationInstrument Startup Getting an Image Moving AroundImage Scanning Zooming In or OutOther Toolbar Buttons System ShutdownImage Options Lower Resolution but quick PhotosPhotography Higher Resolution slower Photos