J E O L J S M - 6 0 6 0 L V S C A N N I N G E L E C T R O N M I C R O S C O P E

6.Close the chamber door by gently pushing it forward until it stops.

7.Press and hold the EVAC button (on the front of the instrument) until it lights up green.

8.WAIT for evacuation of the chamber (about 1 minute). While you’re waiting, go to the next section – Software.

Getting an Image

1.Double click on the SEM Main Menu icon to open the SEM software.

IMPORTANT NOTE: Nearly all of the basic SEM controls may be accessed by clicking on the standard toolbars buttons and the standard display items (buttons above the image, numbers at the bottom of the window, etc.).

2.When the chamber is properly evacuated, the HT button at the top left will be blue and read Ready. If the chamber is vented or still evacuating, the HT button will be grayed out and read Wait.

3.Click on the blue HT-Readybutton to turn on the electron beam. The button should turn green and read HT-ON.

4.Adjust magnification to the lowest possible value (usually between 10X and 20X).

5.Click on the blue AF-Focusbutton on the toolbar or the AUTO FOCUS button on the keypad. This should bring your specimen into focus. If it doesn’t, try turning on the COARSE button and adjusting the focus knob on the keypad.

6.Click on the blue ACB to automatically adjust the contrast and brightness.

7.Click on the blue AS-Stigmabutton for automatic astigmatism correction.

8.You should have a focused image at this point. If so, congratulations…you’re an electron microscopist. If not, call your instructor or teaching assistant for help.

IMPORTANT SETTINGS

oAccelerating Voltage. Check the accelerating voltage value shown at the bottom of the window (Acc. Volt). If you’re happy with your voltage, then continue being happy. If you’d like to change the accelerating voltage, click once on the voltage number, then double click on the desired voltage. A setting of 15 kV works well for most specimens, but feel free experimenting with other accelerating voltages.

oSpot Size. A spot size between 40 and 60 works well for most specimens, but feel free experimenting with the spot size value. You’ll need to adjust the focus, contrast, and brightness after you change the spot size value.

Moving Around

1.The motorized x and y axes are amazing! If you want to move your specimen, simply double- click on the area of interest to re-center the image on this spot, or click and drag the image to move to a spot of interest. It’s kind of like Mapquest.

2.Rotation. Turn the small knob at the bottom of the chamber door assembly (between the x and y axis knobs) to rotate the stage.

8

Image 9
Contents Operating Instructions Table of Contents Introduction SafetyBackground References Sample Holders SamplesSample Preparation Conductive Materials Sample Preparation Non-conductive MaterialsPage Instrument Operation Instrument StartupSample Loading Moving Around Getting an ImageZooming In or Out Image ScanningOther Toolbar Buttons System ShutdownLower Resolution but quick Photos PhotographyImage Options Higher Resolution slower Photos