Emerson MCU200 installation instructions Contents, Specification

Page 1

Installation Leaflet

IP227, Rev. AA

MCU200

July 07

MCU200 Installation, Operation & Maintenance Instructions

Contents

1.Sensor installation

2.Control unit installation

3.Applications: 3.1Gain adjustment 3.2Liquid level alarm

3.3Pump control and latching arms 3.4Blanket or interface detection

4.Spare parts and fault finding

WARNING:

If this equipment is used in a manner not specified by the manufacturer, the protection provided may be impaired. All installation and commissioning of this equipment must be carried out by electrically competent persons

Mobrey ultrasonic liquid level control or sludge blanket or interface detection systems using control unit type MCU200

Each Mobrey ultrasonic control system requires a sensor to suit the specific application, plus a control unit.

These installation instructions cover the Mobrey control units in the MCU200 series.

Specification

Power supply

MCU 201

110/120 ±10%

220/240±10%

50-60 Hz

Installation category

MCU 201

II-IEC664 for 230V ac supply

III-IEC664 for 115V ac supply

Pollution degree

MCU 201

2-IEC664

 

 

 

Power consumption

MCU 201

6VA approx at 240 Vac

 

 

Power supply

MCU 203

24V dc (20V Min, 30V Max) Supply must be floating or negative earth

Current consumption

MCU 203

0.1A Max

 

 

 

Relay output

 

5A at 230V ac DPCO

 

 

Relay state

 

Normal state selectable energised/de-energised

Relay delay

 

0.5,2,8,30 seconds selectable. Operates for change of relay state in

 

 

one direction only. 50msecs in other direction. (Approx.)

LED indicators

 

Red for alarm,

Green for normal,

Yellow for cable fault

LED state

 

Green/Red indication selectable for either sensor state

Sensors compatible

 

Any Mobrey ultrasonic gap, Hisens or Interface sensor

Sensor frequency

 

Switch selects electronics operational at either l MHz or 3.7 MHz

Cable check

 

Option selectable for certain sensors

External input

 

Used to hold relay de-energised to provide pump control

Box size

 

200 x 120 x 75mm

 

 

Fixing centres

 

188 x 188

 

 

 

Box rating

 

IP65 Polycarbonate (clear lid)

 

 

Temperature

 

-40oC to 55oC ambient

 

 

Holes for glands

 

3 off 16mmØ

 

 

 

EMC

 

Emissions: EN50081-1

Immunity: EN50082-1

Safety

 

EN61010-1

 

 

 

 

 

 

 

 

 

www.mobrey.com

Image 1
Contents Specification ContentsSwitching levels and orientation Sensor InstallationInstallation of sensor General descriptionControl Unit Installation Ii Frequency selection Switch Settings in MCU200 Series1MHz sensors Iii Cable check option selectionLevel Alarm Low level alarm, gap type sensor Applications Gain adjustmentHigh level alarm, gap type sensor Special casesPump control and latching alarms High level alarm, Hisens sensorSpecial site conditions Settled sludge discharge control Overflow alarm or fine solids detectionEmerson Process Management Mobrey Measurement Rosemount Inc Interface detector by reflection method