Agilent Technologies N3280A manual 15mA Range Current Measurement Calibration, 101

Page 101

Performance and Calibration Procedures

15mA Range Current Measurement Calibration

Action

Program Commands

1.Jumper the High sense terminal to the High output terminal. Jumper the Low sense terminal to the Low output terminal. Connect the current input of the 3458A multimeter directly to output 1.

2.Select the 15mA range current measurement calibration for output 1. “CAL:CURR:MEAS 0.015,(@1)”

3.

Select the calibration point. *OPC? prevents processing of all

“CAL:LEV P1;*OPC?”

 

subsequent commands to ensure that the output is stable.

 

4.

Set the 3458A multimeter to the 10mA range, measure the output

“CAL:DATA <value>”

 

current, and enter the data into the dc source.

 

Current Priority Mode Programming and 0.5mA Range Measurement Calibration

Action

Program Commands

1.Jumper the High sense terminal to the High output terminal. Jumper the Low sense terminal to the Low output terminal. Connect the current input of the 3458A multimeter directly to output 1.

2.

Select current calibration for output 1.

“CAL:CURR (@1)”

3.

Select the first calibration point. *OPC? prevents processing of all

“CAL:LEV P1;*OPC?”

 

subsequent commands to ensure that the output is stable.

 

4.

Set the 3458A multimeter to the 1mA range, measure the output

“CAL:DATA <value>”

 

current, and enter the data into the dc source.

 

5.

Select the second calibration point. *OPC? prevents processing of

“CAL:LEV P2;*OPC?”

 

all subsequent commands to ensure that the output is stable.

 

6.

Measure the output current and enter the data into the dc source.

“CAL:DATA <value>”

Saving the Calibration Constants

WARNING: Storing calibration constants overwrites the existing ones in non-volatile memory. If you are not sure you want to permanently store the new constants, omit this step. The dc source calibration will then remain unchanged.

 

Action

Program Commands

1.

Save all of the calibration constants.

“CAL:SAVE”

2.

Exit Calibration mode. (*RST also exits calibration mode)

“CAL:STAT OFF”

101

Image 101
Contents Agilent Technologies Model N3280A Component Test DC Source USER’S GuideCertification Warranty InformationGeneral Safety SummarySafety Symbols Declaration Printing History Acoustic Noise InformationTable of Contents Scpi Command Completion Using Device Clear Types of Scpi CommandsExternal References Gpib Capabilities of the DC Source Introduction to ScpiMeasurement Commands Scpi Programming Commands At a Glance Calibration CommandsOutput Commands Status CommandsTrigger Commands System CommandsCommon Commands Voltage Priority TestsTransient Response Tests Error Number List 103Ripple and Noise Tests Page Document Orientation Safety ConsiderationsDescription Options and AccessoriesRemote Programming Option DescriptionVoltage Priority Operation Output CharacteristicsCurrent Priority Operation Output Characteristic Current PriorityMeasurement Characteristics Start of a Measurement Inspection CleaningDamage Packaging Material Additional ItemsBench Operation LocationRack Mounting Do not block the fan exhaust at the rear of the unitConnect the Power Cord Power ConnectionsOutputs 1 Turn the unit off before connecting any wiresOutput Connections Voltage Drops and Lead Resistance Current RatingsAmpacity and Resistance of Stranded Copper Conductors AWG No Maximum Ampacity Resistance at 20 deg. C Free airGuard Connections for Test Fixtures Coaxial Guard ConnectionsComputer Connections External Trigger ConnectionsMaintaining Stability OVP ConsiderationsSwitch Setting Settings for Power Module Configuration SwitchSwitch Setting Address Gpib AddressFront Panel Description Explanation Checkout ProcedureProcedure Runtime Error Messages Selftest Error MessagesPower-On Selftest Errors Case of TroublePage Gpib Capabilities of the DC Source External ReferencesGpib References Scpi ReferencesConventions Used in This Guide Types of Scpi CommandsIntroduction to Scpi Including Common Commands Multiple Commands in a MessageUsing Queries Moving Among SubsystemsChannel List Parameter Types of Scpi MessagesMessage Unit Scpi Data Formats Suffixes and Multipliers Scpi Command CompletionResponse Data Types Class Suffix Unit Unit with MultiplierOutputstate Example Using Device ClearPage Introduction Power-on InitializationProgramming the Output Enabling the OutputOutput Current Output ModeOvervoltage Protection Output Trigger Model Triggering Output ChangesOscillation Protection Selecting the Output Trigger Source Setting the Voltage and Current Trigger LevelsEnabling the Output Trigger System Average Measurements Making MeasurementsGenerating Output Triggers Measurement Samples and Time Interval Power Line CyclesCurrent Ranges Triggered MeasurementsWindow Functions Returning All Measurement Data From the Data BufferSelecting the Measurement Trigger Source Measurement Trigger ModelEnabling the Measurement Trigger System Generating Measurement Triggers Output Settling DelaySelecting the Sensing Function Pre-trigger and Post-trigger Acquisition Pre-trigger and Post-trigger Data AcquisitionDC Source Status Model Programming the Status RegistersOperation Status Group Bit Configurations of Status RegistersStandard Event Status Group Questionable Status GroupStatus Byte Register MSS and RQS BitsServicing Operation Status and Questionable Status Events Determining the Cause of a Service InterruptProgramming Parameters Subsystem CommandsCommon Commands Save new cal constants in non-volatile memory Scpi Programming Commands At a GlanceSubsystem Commands Syntax FUNCtion Mode mode, @list VOLTage Mode mode, @list DELaySets the output voltage in voltage priority mode Sets the triggered output voltage in voltage priority modeESE n Common Commands SyntaxSRE n Output Programming ParametersCALibrateCURRent Calibration CommandsCALibrateCURRentLIMitPOSitive CALibrateCURRentLIMitNEGative CALibrateCURRentMEASureCALibrateDATA CALibratePASSwordCALibrateDATE CALibrateLEVelCALibrateVOLTage CALibrateSAVECALibrateSTATe FETChARRayCURRent? FETChARRayVOLTage? Measurement CommandsFETChCURRent? FETChVOLTage? Query SyntaxMEASureCURRent? MEASureVOLTage? MEASureARRayCURRent? MEASureARRayVOLTage?SENSeCURRentRANGe Enter values greater than 0.015ASENSeSWEepNPLCycles SENSeFUNCtionSENSeSWEepOFFSetPOINts Query Syntax SENSeFUNCtion? @channel listSENSeWINDow SENSeSWEepPOINtsSENSeSWEepTINTerval OUTPut Output CommandsOUTPutOSCProtect OUTPutPROTectionCLEarSOURceCURRentIMMediate SOURceCURRentTRIGgered Default SuffixSOURceCURRentLIMitIMMediate SOURceCURRentLIMitTRIGgered SOURceCURRentLIMitBWIDthSOURceCURRentMODE SOURceCURRentLIMitMODE SOURceDELaySOURceDELayMODE SOURceVOLTageALCBWIDth SOURceFUNCtionMODESOURceVOLTageIMMediate SOURceVOLTageTRIGgered Query Syntax SOURceFUNCMODE? @channel listSOURceVOLTagePROTectionSTATe SOURceVOLTageMODEQuery Syntax SOURceVOLTagePROTectionSTATe? @channel list NR10 orSTATusOPERationENABle Status CommandsBit Configuration of Operation Status Registers STATusOPERationEVENt?Parameters Preset Value STATusPRESetCommand Syntax STATusPRESet Parameters None Examples STATusOPERationNTR STATusOPERationPTRQuery Syntax STATusQUEStionableENABle? @channel list STATusQUEStionableENABleSTATusQUEStionableCONDition? STATusQUEStionableNTR STATusQUEStionablePTRSYSTemVERSion? System CommandsSYSTemERRor? ABORt Trigger CommandsINITiateNAME TRIGgerACQuireTRIGgerTRANsientSOURce TRIGgerACQuireSOURceTRIGgerTRANsient RST ValueCommand Syntax *CLS Parameters None Common CommandsParameters Power-On Value Examples Bit Configuration of Standard Event Status Enable RegisterQuery Syntax *OPT? Returned Parameters Aard Field InformationExample RST Settings Power-on ValueQuery Syntax TST? Returned Parameters NR1 Bit Configuration of Status Byte RegisterTable A-1. Specifications Table A-2 Supplemental Characteristics Safety Output Common ModeDimensions WeightFigure A-1. Output Impedance Graphs all outputs Equipment Required Measurement Techniques Performance & Verification TestsProgramming Test SetupElectronic Load Voltage Priority Tests Program CommandsVoltage Programming and Readback Accuracy Positive Current Limit +CLNegative Current Limit -CL Action Program CommandsCurrent Programming and Readback Accuracy Current Priority TestsVoltage Priority, Constant Voltage Load Effect Load Effect TestsVoltage Priority, -Current Limit Load Effect Test Voltage Priority, +Current Limit Load EffectVoltage Priority, Constant Voltage Source Effect Source Effect TestsCurrent Priority Constant Current Test Voltage Priority, -Current Limit Source Effect Voltage Priority, +Current Limit Source EffectCurrent Priority, Constant Current Source Effect Voltage Priority Ripple and Noise Ripple and Noise TestsCurrent Priority Ripple and Noise Voltage Priority, Transient Recovery TimeTransient Response Tests Figure B-2. Transient Waveform Voltage Priority Current Priority Transient Recovery TimeFigure B-3. Transient Waveform Current Priority Performance Test Equipment FormPerformance Test Record Form Enable Calibration Mode Performing the Calibration ProcedurePositive Current Limit Calibration Negative Current Limit Calibration5A Range Current Measurement Calibration 100101 15mA Range Current Measurement CalibrationSaving the Calibration Constants Calibration Error Messages Changing the Calibration PasswordTable B-3. Gpib Calibration Error Messages Error Meaning103 Error Number ListTable C-1. Error Numbers 104 Error Messages105 Error Messages CPage 107 Line Voltage SelectionPage Mating Connector Part Numbers Earlier Version Output ConnectorsRear Panel Pinout Assignments CHG-2010-J01010-KEPPage 111 Index112 IndexScpi 113114 Canada Australia/New Zealand United States Latin AmericaEurope Asia Pacific JapanManual Updates