Adams AQT SERIES manual Error Messages, Auto Calibration for ALL Other Models

Page 12

8.2AUTO CALIBRATION FOR ALL OTHER MODELS

Before entering calibration mode, press [On/Off] to turn the power on for 1 minute.

Enter calibration mode

Press [On/Off] to turn the power off.

Press [Unit] first and then press [On/Off], while keeping the [Unit] key pressed. The display will show the last calibration value used (For example, “ L – 1000”), press [Tare] to choose between the two calibration weight values. When you have made your selection press [Unit] and the display will show the A/D value.

Calibration

Wait for the stable indicator to be displayed,

Press [Unit] -the display will show "00SAVE". After 2 seconds, the display will return to A/D value.

Place the required weight on the platform. This is the weight that you selected as your calibration weight value earlier.

After the stable indicator is displayed press [Unit]

The display will show "00SAVE". The calibration in complete now.

Press [On/Off] to turn the power off. Press [On/Off] to turn on the power again,

Place full weight (eg.2000g for AQT2000) on the platform to verify weighing is correct. If not, repeat the calibration procedure.

9.0ERROR MESSAGES

EEEEHI

Overload

 

 

 

 

EEEELO

A/D value too low

 

 

 

 

 

 

 

Low battery

 

 

 

 

 

 

 

Note: Calibration may be required before weighing. Read section 8.0 on "CALIBRATION" first and if necessary calibrate your balance for accurate weighing.

Image 12
Contents AQT Series Contents Introduction Specifications Model # AQT-200 AQT-600 AQT-1500 AQT-2000 AQT-5000Safety Precautions Parts Description TOPSetting UP the Balance Installing Batteries and PowerPreparation Weighing Procedures WeighingBefore Weighing Setting of Functions Enter Function Setting ModeSelection of Auto Shut OFF Mode Selection of Auto Backlight ModeUnits Selection for AQT 200 only Calibration Auto Calibration for AQT 200 onlyError Messages Auto Calibration for ALL Other ModelsReplacement Parts and Accessories Service InformationWarranty Information FCC Compliance Adam Equipment Company