Polaris D7500 manual Maintenance, Inspection and Cleaning

Page 21

MAINTENANCE

INSPECTION AND CLEANING

Perform the following inspection and cleaning procedures periodically as a preventive maintenance measure to keep the Polaris System in optimal condition resulting in trouble-free operation producing crisp, clear images..

Do not attempt any internal service of Polaris System components. Contact your local autho- rized Air Techniques dealer for service. Failure to heed this warning may result in equipment damage and voiding the warranty.

Inspection - Routinely inspect each component of the Polaris System for possible defects as follows:

1..Camera Handpiece -

a..Check overall handpiece for chips, cracks or other irregularities..

b..Check the lens window for debris or spots..

c..Check the connector socket for damage..

2..Control Module -

a..Check for broken indicator lamp and switch..

b..Check all connector sockets for damage, loose or missing hardware..

3..Cables and Connectors -

a..Check cables for damaged or deteriorated insulation kinking or twisting..

b..Check connectors for loose, bent or missing pins..

c..Check that both quick-disconnect plug ends snap closed to snugly secure the Handpiece Cable to the Camera Handpiece and Control Module connectors..

CAUTION: Always turn Off power and unplug the Polaris System components before performing cleaning procedures.

Do not spray cleaning liquids or disinfectants directly on the camera handpiece. Use care not to allow liquids to run into internal circuitry.

Do not wipe the surfaces using benzine, thinner, etc. as this may degrade the finish. Do not autoclave the Camera Handpiece.

2The camera cover must be used for only one patient and disposed of properly in accordance with local code.

Cleaning - Clean each component of the Polaris System as needed by performing the following procedures..

1..Control Module-Use a non-corrosive surface disinfectant such as Birex® se..

2..Camera Handpiece-Following every patient use, remove the camera cover from the Camera Handpiece and dispose in accordance with local regula- tions. The Camera Handpiece can be wiped clean and disinfected with any non-corrosive surface disinfectant such as Birex® se..

3..Camera Window Lens-Remove debris or spots from lens window by cleaning with alcohol and cotton swabs.

21

Image 21
Contents I v e r s a l Foreword Table of ContentsUse of Accessory Equipment General SafetyDo Not Attempt Internal Service ContraindicationsMarkings CongratulationsImportant Information General NotesIncorrect Usage BOX Contents KEY Parts Identification Component DescriptionSupplied Polaris System Components Component Description PRE-INSTALLATION Information System PropertiesGeneral Tab Polaris Camera Handpiece Handpiece FunctionsControl Module Front Panel Controls and Indicators Control Module Rear Panel ConnectorsControl Module Functions Installation System InterconnectionsSystem Interconnection Handpiece Holder Chair-side Adapter Assembly By turning the Joint Connector Hex Bolt 4 clockwiseCamera Handpiece Connection Control Module ConnectionUSB Cable Connection Directly to PC Optional Video Connections Operating Power ConnectionPolaris Handpiece Storing Camera PortabilityPRE-OPERATION Camera Cover InstallationActivation System OperationComputer-Connected System Operation The Polaris System is System Operation Tab Opens Video Proc Amp Screen Diagnostic ProceduresInspection and Cleaning MaintenancePhysical Characteristics Specifications10 to 40º C 50 to 104º F To 70º C 32 to 158º FPolaris Camera System AccessoriesVisix Imaging Software Key FeaturesON-LINE Warranty Registration WarrantyPage Page  Stsj