LOW REFLECTION CD TRAVERSE ADJUSTMENT | CD TRAVERSE ADJUSTMENT | ||||||||||||||||||||
Connection : | Connection : | ||||||||||||||||||||
|
| oscilloscope |
| oscilloscope | |||||||||||||||||
MAIN board |
|
|
|
|
|
|
|
| MAIN board |
|
|
|
|
|
|
|
| ||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||||||
SP526 (TE) |
|
|
|
|
|
|
|
|
| SP526 (TE) |
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||
AP503 (VC) |
|
|
|
|
|
|
|
|
| AP503 (VC) |
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
| ||||||||||||
|
|
|
|
|
|
|
|
|
|
Adjustment Method :
1.Set the servo mode of the test mode (Mode : 000).
2.Press the ( key, and set the low reflection CD test adjustment mode (Mode : 040) using the VOLUME + or – key.
3.Insert any MO disk available on the market.
4.When the ( key is pressed, the low reflection CD EF balance adjustment mode (Mode : 042) will be set after low reflection CD focus search adjustment mode (Mode : 041).
5.Press the P key to perform automatic adjustment, and check that the traverse waveform is symmetrical at the top and bottom.
(Traverse waveform)
A
0 V |
|
| C |
|
| ||
|
| ||
|
|
|
|
B
Specification : A = B, C >= 0.9
6.Check that the traverse level at this time is above 0.9
7.Press the pkey.
8.Exit the test mode.
Adjustment Location : See page 19.
Adjustment Method :
1.Set the servo mode of the test mode (Mode : 000).
2.Press the ( key, and set the CD test adjustment mode (Mode : 050) using the VOLUME + or – key.
3.Press the =or + key and move the optical
4.Insert a test CD
5.When the ( key is pressed, the CD EF balance adjustment mode (Mode : 052) will be set after CD focus search adjustment mode (Mode : 051).
6.Press the P key to perform automatic adjustment, and check that the traverse waveform is symmetrical at the top and bottom.
(Traverse waveform)
A
0 V |
|
| C |
|
| ||
|
| ||
|
| ||
|
|
|
|
B
Specification : A = B, C >= 1.0
7.Check that the traverse level at this time is above 1.0
8.Press the pkey.
9.Exit the test mode.
Adjustment Location : See page 19.
– 18 –