Fluke 45 user manual IEEE-488 Interface Function Subsets, IEEE-488 Common Commands

Page 76

45

Users Manual

Table 5-8. IEEE-488 Interface Function Subsets

SH1

Source Handshake

AH1

Acceptor Handshake

T5

Talker

L4

Listener

SRI

Service Request

RL1

Remote/Local

DC1

Device Clear*

DT1

Device Trigger

E1

Electrical interface

^C (CNTRL C) is the RS-232 equivalent of DC1, causing <CR><LF> and a new prompt to be output.

Table 5-9. IEEE-488 Common Commands

Command

*CLS

*ESE <value>

*ESE?

*ESR?

*IDN

*OPC

*OPC?

*RST

Name

Clear Status

Event Status Enable

Event Status Enable Query

Event Status

Register Query

Identification Query

Operation Complete Command

Operation Complete Query

Reset

Description

Clears all event registers summarized in the status byte, except for "Message Available," which is cleared only if *CLS is the first message in the command line.

Sets "Event Status Enable Register" to <value>, an integer between 0 and 255. <value> is an integer whose binary equivalent corresponds to the state (1 or 0) of bits in the register. If <value> is not between 0 and 255, an Execution Error is generated.

EXAMPLE: decimal 16 converts to binary 00010000. Sets bit 4 (EXE) in ESE to 1.

Meter returns the <value> of the "Event Status Enable Register" set by the *ESE command. <value> is an integer whose binary equivalent corresponds to the state (1 or 0) of bits in the register.

Meter returns the <value> of the "Event Status Register" and then clears it. <value> is an integer whose binary equivalent corresponds to the state (1 or 0) of bits in the register.

Meter returns the identification code of the meter as four fields separated by commas. These fields are: manufacturer ("FLUKE"); model ("45"); seven-digit serial number; version of main software and version of display software.

Meter sets the Operation Complete bit in the Standard Event Status Register when parsed.

Meter places an ASCII "1 " In The output Queue when parsed.

Meter performs power-up reset except that the state of IEEE- 488 interface is unchanged, including: 1) instrument address, 2) Status Byte and, 3) Event Status Register.

5-20

Image 76
Contents Dual Display Multimeter Limited Warranty & Limitation of Liability Declaration of the Manufacturer or Importer Table of Contents Applications Appendices List of Tables Page List of Figures Introduction Introducing the Fluke 45 Dual Display MultimeterWhere to go from Here Options and AccessoriesModel Accessories Description Maintenance Appendices Page Multimeter Safety Introduction Getting StartedFront Panel Using the Pushbuttons Turning the Meter onSummary of Basic Pushbutton Operations Automatic Input Terminal Selection Selecting a Measurement RangeTaking Some Basic Measurements Measuring Voltage, Resistance, or Frequency Diode/Continuity TestingMeasuring Current Freq Continuity Testing Operating the Meter Under Battery Power Optional Rack MountingPage Operating the Meter From the Front Panel Front Panel OperationsPrimary Display DisplaySecondary Display Remote SMF Input Terminals Input TerminalsSelecting a Measurement Function RangingAutoranging Input LimitsFunction Buttons Slow Reading Rate Range Full Scale Ohms Ranges and Full Scale Values Fast Reading RateMedium Reading Rate Range Full Scale Frequency Ranges and Full Scale Values Slow and MediumManual Ranging Measuring FrequencyFrequency Ranging Frequency Measurement Rates Measuring Frequency of Current 100 mA and 10A InputsFrequency Sensitivity Selection Frequency Measurement RatesMaximum Sinewave Inputs for Frequency Measurements Range Selecting a Function ModifierMaximim Input Voltage REL Relative Readings Modifier Function Modifier Selection ButtonsDB Decibels and Audio Power Modifier Reference Impedances in OhmsHold Touch Hold Modifier Using Function Modifiers in CombinationMN MX Minimum Maximum Modifier Using the SButton Selecting a Measurement Rate RatePushbuttons 10. SButton OperationsUsing The Compare Comp Function List and Number EditorsUsing the List Editor Using the Number Editor Factory Settings of Power-Up Configuration Power-Up ConfigurationChanging the Power-Up Configuration Configurations 13. Power-Up Configuration Set at Factory ParameterCalibration Using the Dual Display ApplicationsUsing Measurement Functions in Combination Dual Display Showing Volts AC and FrequencySample Dual Display Applications Primary Display ApplicationsDC Voltage and DC Current Measurement on Input Signal Response TimesHow the Meter Makes Dual Display Measurements Slow Rate Auto Single Range1 Range2 Medium Rate Fast Rate Slow Med FastUpdate Rate in the Dual Display Mode External TriggerThermal Voltages AllWhen Measuring Resistance Two-Wire ConfigurationTrue RMS Measurements Correcting for Test Lead ResistanceCalculated AC + DC RMS Measurements Effects of Internal Noise in AC MeasurementsWaveform Comparison True RMS vs. Average-Responding Meters True RMS Measurements Page Local and Remote Operations Computer InterfacesSetting Communication Parameters RS-232 Factory Settings of RS-232 Communication ParametersFactory Setting Preparing the Meter for Operations via the RS-232 InterfaceRS-232 Print-Only Mode Device Clear Using C Cntrl C Cabling the Meter to a Host or Printer RS-232Character Echoing and Deletion RS-232 PromptsInstalling the IEEE-488 Interface Preparing the Meter to be Operated via IEEE-488 InterfaceIEEE-488 Operating Limitations Enabling the IEEE-488 InterfaceGetting Started With An Installation Test Installation Test for RS-232 OperationsInstallation Test for IEEE-488 Operations Cabling the Meter to a HostHow the Meter Processes Input If Test FailsInput Strings Input TerminatorsTypical IEEE-488 Input Strings Sending Command Strings to the MeterSending Numeric Values to the Meter Page Typical IEEE-488 Input Strings Triggering Output How the Meter Processes OutputExternal Triggering from the Front Panel Setting the Trigger Type Configuration External Triggering via the Computer InterfaceService Requests IEEE-488 Only and Status Registers Settling DelayTrigger Type Rear TriggerRead Command Write CommandEnable Register Status Register SummaryESB MAV 3 2 1 Event Status and Event Status Enable Registers Event Status and Event Status Enable RegistersStatus Byte Register Reading the Status Byte RegisterDescription of Bits in ESR and ESE Bit No Name True Set to 1 ConditionsDescription of Bits in the Status Byte Register Bit No Name Service Request Enable RegisterTrue Set to 1 Condition Computer Interface Command Set IEEE-488 Capabilities and Common CommandsIEEE-488 Common Commands IEEE-488 Interface Function Subsets10. IEEE-488 Common Commands Number StateFunction Modifier Commands and Queries Function Commands and Queries11. Function Commands and Queries Primary Secondary Display Command 12. Function Modifier Commands and Queries Description12A. Reference Impedance Values Ref Impedance Value Ref ImpedanceMIN Range and Measurement Rate Commands and Queries Ohms RangeCurrent Range Frequency RangeMeasurement Queries 13. Range and Measurement Rate Commands and Queries14. Measurement Queries Description Compare Commands and Queries Trigger Configuration Commands15. Compare Commands and Queries Description 16. Trigger Configuration Commands DescriptionMiscellaneous Commands and Queries 17. Miscellaneous Commands and Queries Description16A. Measurement Units Output with Format Measurement FunctionRS-232 Remote/Local Configurations 18. Remote/Local Configuration Commands DescriptionSample Program Using the RS-232 Computer Interface Sample Program for RS-232 Computer InterfaceSample Programs Using the IEEE-488 Computer Interface Sample Programs for IEEE-488 Computer InterfaceAam25s.tif Aam26s.tif Aam27s.tif Operating the Meter Using the Computer Interface Introduction Cleaning Line FuseTesting Current Input Fuses Current Input FusesReplacing the 10 a Input Fuse Replacing the 100 mA Input FuseFront Panel Input Terminal Error No Self-Test Error Codes Meaning Self-Test Diagnostics and Error CodesPerformance Tests Service Input LevelFrequency Min MaxUsing decades 300 Ω Short 30 kΩ Replacement Parts Replacement PartsPerformance Tests for mA Current Functions Min Max 29.982 30.018TM1 TM2 TM3 TM4 TM5 Appendices Page RS-232 and IEEE-488 Reading Transfer Rates Display Counts and Reading RatesResponse Times Normal Mode Rejection Ratio Common Mode Rejection RatioDC Voltage Input ImpedanceDecibel Resolution True RMS AC Voltage, AC-CoupledMaximum Input AccuracyDecibel Reference Resistance Maximum Crest Factor Common Mode Rejection RatioAC + DC Voltage Accuracy DC Current AC CurrentRange Slow AccuracyMaximum Crest Factor Maximum Input RangeFrequency Slow Accuracy Medium FastOhms Diode Test/ContinuitySensitivity of AC Voltage FrequencySensitivity Level of AC Current Environmental General OptionsASCII/IEEE-488 Bus Codes Page ASCII/IEEE-488 Bus Codes B IEEE-488.2 Device Documentation Requirements Implementation of Ieee StandardNumeric Response Data Measurement *RDT and *RAT? commands are not implemented
Related manuals
Manual 122 pages 22.93 Kb

45 specifications

The Fluke 45 is a versatile and high-precision Digital Multimeter that is designed for a wide range of applications in electrical testing, troubleshooting, and maintenance. Known for its reliability and advanced features, the Fluke 45 is an essential tool for engineers, technicians, and professionals who require accurate measurement capabilities.

One of the standout features of the Fluke 45 is its ability to measure both AC and DC voltage and current. With a maximum input voltage of 1000 volts and current ranges up to 10 amps, users can conduct a variety of electrical tests with confidence. The device is equipped with an intuitive dual display, allowing for simultaneous viewing of multiple measurements. This feature enhances usability during complex diagnostics.

The Fluke 45 also boasts an impressive resolution of 0.1 mV and a basic accuracy of 0.25%, which translates to highly reliable readings. The auto-ranging capability simplifies the measurement process, automatically selecting the appropriate range based on the input signal, thus saving time and reducing user error.

In addition to voltage and current measurements, the Fluke 45 features resistance measurement capabilities, with a range of up to 40 MΩ. This makes it an effective tool for testing circuits, identifying faults, and ensuring the integrity of electrical systems. The continuity test function, complete with a built-in beeper, assists in quickly diagnosing wiring issues.

Another remarkable characteristic of the Fluke 45 is its frequency measurement capabilities, which can measure frequencies up to 1 MHz. This is particularly useful in applications involving signal analysis and troubleshooting oscillators.

Moreover, the Fluke 45 is designed with user convenience in mind. It is lightweight and portable, making it easy to carry around, whether on a job site or in a workshop. The rugged construction, reinforced with high-quality materials, ensures it can withstand harsh conditions often encountered in the field.

To enhance functionality, the Fluke 45 features a data hold function that freezes the displayed measurement for easier reporting. Additionally, it can store data for later retrieval, making it convenient for users who need to compile test results over time.

In summary, the Fluke 45 Digital Multimeter is an advanced instrument that combines precision, versatility, and user-friendly design. Its wide measurement capabilities, high accuracy, and durable build make it an invaluable tool for professionals in diverse electrical and electronic applications. Whether for routine maintenance or complex troubleshooting, the Fluke 45 stands out as a reliable partner in achieving accurate and efficient results.